PERFUSION REPORTS

REPORT
110
DATE
31/1/2023
TITLE
DISPOSABLE FAULT
DETAILS OF INCIDENT

    During the CPB, it was noticed  that the patient flow (as measured by the Spectrum M4 flow probe) was  significantly reduced (3.9 Lpm) comparative to pump flow at 6.0 Lpm. In  addition, the Spectrum emboli counter was increasing incrementally. At  first, this was thought to be due to flow probe positioning, but with further  investigation transpires that a kink had developed at the reservoir  outlet. 
 
 We concluded that the phenomena experienced was due to cavitation caused by  this kink, possibly resulting in mal-perfusion and/or micro-emboli. No injury  was caused to the patient.  

DETAILS OF INJURY TO PATIENT

No Injury

TYPE OF DEVICE/MANUFACTURER
ACTION TAKEN

A sling was used to ensure no further kinking of the tubing.

RESPONSE FROM MANUFACTURER

COMPLAINT INFORMATION

Complaint #:

2023-00505

Date of Event:

Jan 24, 2023

Hospital:

ABERDEEN ROYAL INFIRMARY

Event Description:

The pump boot outlet kink was noticed during aseptic assembly. An unsuccessful attempt was made to correct this defect, resulting in cutting out of this section of tubing (no pictures taken). The circuit was then deemed adequate for clinical use and CPB was commenced. During the CPB, it was noticed that the patient flow (as measured by the Spectrum M4 flow probe) was significantly reduced (3.9 Lpm) comparative to pump flow at 6.0 Lpm. In addition, the Spectrum emboli counter was increasing incrementally. At first, this was thought to be due to flow probe positioning, but with further investigation transpires that an additional kink had developed at the reservoir outlet (see picture attached). We concluded that the phenomena experienced was due to cavitation caused by this kink, possibly resulting in mal-perfusion and/or micro-emboli. This incident has been reported via the hospital adverse events system (DATIX) and I assume the MHRA will be informed. Luckily, no injury was caused to the patient.

PRODUCT INFORMATION:

Product Number:

IN01728H

Product Descripti on:

ABERDEEN AD INSPIRE 8F

Serial/LOT Number

: 2210270110

Investi gati on Summary:

LivaNova received a report stating that during set up, the Customer noted that the section of tubing #0191061 SG XS with a length of 60cm corresponding to the reservoir outlet in the pump loop was kinked. 1 unit of PTS PN IN01728H, lot 2210270110. No patient/user affected. Pictures of thissue were provided by the customer, confirming the kink. The sample was not available for investigation. Manufacturing records have been verified to be conforming to specifications. From complaints database review, three further similar events have been recorded on this PTS pack code (See #2022-00391, #2022-01109 and #2022-06709). No other complaint was found reported for the concerned batch of 24 units of this lot. Based on the above, the root cause could betraced back to two factors: - a tie-band that pressed against the tubing in question, which during sterilization caused a permanent mark or deformation. - a tubing length selected by the customer which should be changed in a following revision, since it is not long enough, this generates a tension in the tubing during set-up and the procedure, which can generate the kink due to the initial material deformation caused by the tie bands pressing on the tubing. As mitigation action, the drawing of this involved PTS will be updated to revision no. 03 in March 2023 in order to add specific pictures to help the operator in the assembly phase of this involved line in the table pack. The aim of this action is to reduce operator error as far as possible. Therefore, all lots manufactured before the new drawing revision release, are pre-mitigation action implementation. LivaNova maintains and documents periodic customer events monitoring process in order to evaluate actions for products improvement.

Root Causes:

Based on the above, the root cause could be traced back to two factors: - a tie-band that pressed against the tubing in question, which during sterilization caused a permanent mark or deformation. - a tubing length selected by the customer which should be changed in a following revision, since it is not long enough, this generates a tension in the tubing during set-up and the procedure, which can generate the kink due to the initial material deformation caused by the ti e bands pressing on the tubing.

Corrective Actions:

As mitigation action, the drawing of this involved PTS will be updated to revision no. 03 in March 2023 in order to add specific pictures to help the operator in the assembly phase of this involved line in the table pack. The aim of this action is to reduce operator error as far as possible. Therefore, all lots manufactured before the new drawing revision release, are pre-mitigation action implementation. LivaNova maintains and documents periodic customer events monitoring process in order to evaluate actions for products improvement.

Best regards,

Customer Quality-Team

Complaints Handling # 2023-00505

Kind regards,

Marti n

Marti n Brunt

Territory Sales Manager (CP & ATS)

Cardiac Surgery Division

United Kingdom

ADDTIONAL INFORMATION

It is interesting to note that as we monitor actual flow in the tube against the flow calculated by the roller pump it can show up this sort of problem highlighting the importance of inline monitoring.

 As the kink is from a pre-connected circuit that has suffered a imposed memory in the packaging and deformed on bypass.

Tubing length has now been altered to mitigate this problem.

SUPPORTING VISUALS/VIDEO
SUPPORTING DOCUMENT

OTHER PERFUSION
 REPORTS

London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Select and scroll to view chosen report
REPORT NO
089
1/9/21
Sechrist Gas Blender Sweep on ECMO
REPORT NO
088
1/9/21
Oxygenator High Pressure Events
REPORT NO
087
1/10/21
Leaking one way valve
REPORT NO
086
1/7/21
Electronic Gas Blender Fault
REPORT NO
085
1/2/21
Protamine Administration Issue
REPORT NO
084
1/6/21
Damaged roller pump lid
REPORT NO
083
1/6/21
IABP malfunction
REPORT NO
082
1/1/21
Accidental cardioplegia administration
REPORT NO
081
1/4/21
S5 Double Roller Pump incident
REPORT NO
080
1/5/21
Human factors issue
REPORT NO
079
1/6/21
Centrimag battery module failure
REPORT NO
078
1/6/21
Sampling manifold valve fault
REPORT NO
077
1/5/21
IABP transport issue
REPORT NO
076
1/4/21
Modified Ultrafiltration (MUF) Issue
REPORT NO
075
1/3/21
Electromagnetic interference with HLM
REPORT NO
074
1/2/21
HCU40 Failure
REPORT NO
073
1/2/21
HCU40 valve failures
REPORT NO
072
1/1/21
HCU40 failure
REPORT NO
071
1/1/21
LIVA NOVA STATEMENT ON CENTRIFUGAL HEAD
REPORT NO
070
1/1/21
Loose component on circuit
REPORT NO
069
1/12/20
Centrifugal Head noise
REPORT NO
068
1/11/20
Venous Reservoir disposable issue
REPORT NO
067
1/11/20
Failure of Heater Cooler due to leaking valves
REPORT NO
066
1/11/20
Failure of Valves on heater cooler unit
REPORT NO
065
1/9/20
Intra Aortic Balloon Pump (IABP) Failure
REPORT NO
064
1/9/20
Centrifugal head failure
REPORT NO
063
1/8/20
Centrifugal head issue 4
REPORT NO
062
1/8/20
Human Factors Issue
REPORT NO
061
1/8/20
LIVA NOVA RESPONSE TO REPORTED CENTRIFUGAL HEAD ISSUE
REPORT NO
060
1/8/20
Oxygenator Fibre Leak
Perfusion Report cards are shown in batches of 30 at any one time where available.
120
119
118
117
116
115
114
112
111
110
109
108
107
106
105
104
103
102
101
100
099
098
097
096
095
094
093
092
091
090
089
088
087
086
085
084
083
082
081
080
079
078
077
076
075
074
073
072
071
070
069
068
067
066
065
064
063
062
061
060
059
058
057
056
055
054
053
052
051
050
049
048
047
046
045
044
043
042
041
040
039
038
037
036
035
034
033
032
031
030
029
028
027
026
025
024
023
022
021
020
London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Know the report number? Scroll horizontally through the circle icons and select to view the appropriate report.