FSN REPORTS

FSN
006
DATE
1/1/2014
TITLE
Maquet Cardioplegia Temperature Reading Inaccuracies
PROBLEM

Maquet Cardiovascular have issued a Safety Notice concerning the temperature reading on a certain number of PLEGIOX Cardioplegia heat exchangers. The problem is limited to selected lot numbers and only affects a small number of hospitals in the UK. The cause of the problem is a change to a moulded component, which restricts the depth the external temperature probe can be inserted into the device. There could be up to a three degree Celsius difference in read and actual temperature however the performance of the heat exchanger is unaffected.

ACTION TAKEN

The Safety Committee would like to make the UK community aware of this situation, which is being dealt with by the Maquet Technical Services team. Further information and affected lot numbers can be found in the following documents Plegiox Field Safety Notice.pdf‎  and Plegiox Lot Numbers.pdf, which can be accessed below:

DOWNLOADS
USEFUL LINKS

OTHER FSN REPORTS

London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Select and scroll to view FSN Details
REPORT NO
016
1/8/2020
Getinge HCU40 Replacment of Vacuum Valves
REPORT NO
015
1/7/2020
Getinge Replacement IFU's for HU35 and rotaflow console
REPORT NO
014
1/7/2020
Liva Nova Perceval Valve Accessory kit
REPORT NO
013
1/7/2020
Medtronic HeartWare HVAD system lubricant
REPORT NO
012
1/6/2020
Getinge Quadrox Holder FSN
REPORT NO
011
1/5/2020
Teleflex IABP
REPORT NO
010
1/4/2020
Medtronic HeartWare HVAD Systems
REPORT NO
009
1/3/2017
Infections Associated With HCUs In CPB & ECMO
REPORT NO
008
1/9/2015
Potential Mycobacterial Contamination Of HCUs
REPORT NO
007
1/2/2014
SWISSMEDIC: Water-Borne Infection In Implant Patients
REPORT NO
006
1/1/2014
Maquet Cardioplegia Temperature Reading Inaccuracies
REPORT NO
005
1/1/2012
Maquet Adhesive Failure On Oxygenator Blood Outlet Connector
REPORT NO
004
1/8/2011
Sorin Heat Exchanger Blood Water Leak
REPORT NO
003
1/6/2011
Maquet Transient Increase In Pump Speeds
REPORT NO
002
1/3/2011
Terumo CDI500 Possibility Of Parameter Drift
REPORT NO
001
1/2/2009
Datascope IABP Harness Not Fit For Purpose
034
033
032
031
030
029
028
027
026
025
024
023
022
021
020
019
018
017
016
015
014
013
012
011
010
009
008
007
006
005
004
003
002
001
London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Know the report number? Scroll horizontally through the circle icons and select to view the appropriate report.