TITLE
SWISSMEDIC: Water-Borne Infection In Implant Patients
PROBLEM

We have been advised by the MHRA that Swissmedic, the Swiss Agency for Therapeutic Products has recently published a safety notice written by the Federal Office of Public Health (FOPH) in Switzerland concerning a rare number of cases of a water borne infection in cardiac surgery patients that have had implants. The effects of such a bacterial infection can take one or two years to develop in some cases.They have had 6 known cases between 2008 and 2012.

With regards to the cause of these infections the FOPH report states:

"The cause of the infection is suspected to be the contamination of a technical device (heater cooler unit) that is used in hospitals during open heart surgery to control the temperature of the blood. These devices are operated with filtered tap water. The bacterium was detected in the water and in the discharged air of the devices in several Swiss hospitals. It is possible that during very few of these operations, a transmission of the bacterium occurred. However, the exact route of transmission has not yet been established."

Whilst it is not conclusive that the heat-cooler is the cause of these infections the Safety Committee felt it helpful to remind the membership of the importance of following the manufacturers guidelines on disinfecting and cleaning their heater-coolers and ensuring the departmental protocols include these practices.

Downloads: Sorin guidelines for the disinfecting and cleaning their heater-coolers.

ACTION TAKEN

N/A

DOWNLOADS
USEFUL LINKS

OTHER FSN REPORTS

London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Select and scroll to view FSN Details
REPORT NO
034
28/2/2022
Medtronic HVAD
REPORT NO
033
1/10/2021
Gettinge Custom tubing notification
REPORT NO
032
1/9/2021
Getinge Cardiosave Battery FSN
REPORT NO
031
1/8/2021
Getinge HCU40
REPORT NO
030
1/7/2021
Getinge Tubing FSN
REPORT NO
029
1/6/2021
Medtronic HVAD Update June 2021
REPORT NO
028
1/5/2021
Getinge Sterile Barrier Integrity issue FSN
REPORT NO
027
1/5/2021
Terumo Extracorporeal Circuit Integrity
REPORT NO
026
1/3/2021
Datascope IABP Cypersecurity vulnerabilities
REPORT NO
025
1/5/2021
Update to Medtronic HVAD from Dec 2020
REPORT NO
024
1/2/2021
FSN Medtronic Pixie
REPORT NO
023
1/2/2021
Getinge IABP Battery Usage, Charging, Storage
REPORT NO
022
1/2/2021
Getinge HCU40 Replacement of Vacuum Valves
REPORT NO
021
1/12/2020
Getinge Paediatric Hardshell Reservoir
REPORT NO
020
1/11/2020
Getinge HLS FSN
REPORT NO
019
1/10/2020
Getinge HCU40 cable connection FSN
REPORT NO
018
1/9/2020
Getinge BMU40 FSN
REPORT NO
017
1/8/2020
Getinge Device Recall IAB
034
033
032
031
030
029
028
027
026
025
024
023
022
021
020
019
018
017
016
015
014
013
012
011
010
009
008
007
006
005
004
003
002
001
London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Know the report number? Scroll horizontally through the circle icons and select to view the appropriate report.