FSN REPORTS

FSN
007
DATE
1/2/2014
TITLE
SWISSMEDIC: Water-Borne Infection In Implant Patients
PROBLEM

We have been advised by the MHRA that Swissmedic, the Swiss Agency for Therapeutic Products has recently published a safety notice written by the Federal Office of Public Health (FOPH) in Switzerland concerning a rare number of cases of a water borne infection in cardiac surgery patients that have had implants. The effects of such a bacterial infection can take one or two years to develop in some cases.They have had 6 known cases between 2008 and 2012.

With regards to the cause of these infections the FOPH report states:

"The cause of the infection is suspected to be the contamination of a technical device (heater cooler unit) that is used in hospitals during open heart surgery to control the temperature of the blood. These devices are operated with filtered tap water. The bacterium was detected in the water and in the discharged air of the devices in several Swiss hospitals. It is possible that during very few of these operations, a transmission of the bacterium occurred. However, the exact route of transmission has not yet been established."

Whilst it is not conclusive that the heat-cooler is the cause of these infections the Safety Committee felt it helpful to remind the membership of the importance of following the manufacturers guidelines on disinfecting and cleaning their heater-coolers and ensuring the departmental protocols include these practices.

Downloads: Sorin guidelines for the disinfecting and cleaning their heater-coolers.

ACTION TAKEN

N/A

DOWNLOADS
USEFUL LINKS

OTHER FSN REPORTS

London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Select and scroll to view FSN Details
REPORT NO
016
1/8/2020
Getinge HCU40 Replacment of Vacuum Valves
REPORT NO
015
1/7/2020
Getinge Replacement IFU's for HU35 and rotaflow console
REPORT NO
014
1/7/2020
Liva Nova Perceval Valve Accessory kit
REPORT NO
013
1/7/2020
Medtronic HeartWare HVAD system lubricant
REPORT NO
012
1/6/2020
Getinge Quadrox Holder FSN
REPORT NO
011
1/5/2020
Teleflex IABP
REPORT NO
010
1/4/2020
Medtronic HeartWare HVAD Systems
REPORT NO
009
1/3/2017
Infections Associated With HCUs In CPB & ECMO
REPORT NO
008
1/9/2015
Potential Mycobacterial Contamination Of HCUs
REPORT NO
007
1/2/2014
SWISSMEDIC: Water-Borne Infection In Implant Patients
REPORT NO
006
1/1/2014
Maquet Cardioplegia Temperature Reading Inaccuracies
REPORT NO
005
1/1/2012
Maquet Adhesive Failure On Oxygenator Blood Outlet Connector
REPORT NO
004
1/8/2011
Sorin Heat Exchanger Blood Water Leak
REPORT NO
003
1/6/2011
Maquet Transient Increase In Pump Speeds
REPORT NO
002
1/3/2011
Terumo CDI500 Possibility Of Parameter Drift
REPORT NO
001
1/2/2009
Datascope IABP Harness Not Fit For Purpose
034
033
032
031
030
029
028
027
026
025
024
023
022
021
020
019
018
017
016
015
014
013
012
011
010
009
008
007
006
005
004
003
002
001
London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Know the report number? Scroll horizontally through the circle icons and select to view the appropriate report.