PERFUSION REPORTS

REPORT
002
DATE
1/1/2010
TITLE
Luer Cap Found In Venous Inlet Port
DETAILS OF INCIDENT

After instigating cardiopulmonary bypass for an urgent procedure, it was noticed the venous return was not satisfactory to sustain full flow despite surgical manipulation. CPB was discontinued, the venous line disconnected at the inlet of the venous reservoir where a non-hospital standard luer cap was found lodged inside the venous inlet port. This was removed, CPB re-instituted and no further problems were encountered. No harm came to the patient.

DETAILS OF INJURY TO PATIENT
TYPE OF DEVICE/MANUFACTURER
ACTION TAKEN

The Safety Committee recommends that a thorough inspection of all blood path ports be made during circuit assembly and suitable flow and pressure checks be performed prior to establishing CPB. Consider incorporating into departmental checklists if not already included.

RESPONSE FROM MANUFACTURER
ADDTIONAL INFORMATION
SUPPORTING VISUALS/VIDEO
SUPPORTING DOCUMENT

OTHER PERFUSION
 REPORTS

London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Select and scroll to view chosen report
REPORT NO
002
1/1/10
Luer Cap Found In Venous Inlet Port
REPORT NO
001
1/1/09
Oxygenator Water To Blood Leak
Perfusion Report cards are shown in batches of 30 at any one time where available.
123
122
121
120
119
118
117
116
115
114
112
111
110
109
108
107
106
105
104
103
102
101
100
099
098
097
096
095
094
093
092
091
090
089
088
087
086
085
084
083
082
081
080
079
078
077
076
075
074
073
072
071
070
069
068
067
066
065
064
063
062
061
060
059
058
057
056
055
054
053
052
051
050
049
048
047
046
045
044
043
042
041
040
039
038
037
036
035
034
033
032
031
030
029
028
027
026
025
024
023
London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Know the report number? Scroll horizontally through the circle icons and select to view the appropriate report.