PERFUSION REPORTS

REPORT
101
DATE
6/4/2021
TITLE
Irregular blood flow/foreign looking substance in Oxygenator
DETAILS OF INCIDENT

Liva Nova Inspire 6F  (integrated arterial filter)  

Elective  CABG patient with standard medical history & no known coagulation pathologies. Full systemic anticoagulation was achieved & confirmed pre-CPB with an ACT over 450 seconds.

Immediately following the onset of bypass, the arterial filter was observed to have turbulent flow & higher than expected RPM's for the generated flows. The blood flow in the filter appeared to be turbulent and there was a gelatinous like appearance in the arterial line filter.

Visually all looked fine and arterial blood looked oxygenated. ACT, arterial & venous blood gases were sent immediately & indicated that there no issues with oxygenation  & CO2 removal.

There was no signs of this material elsewhere within the circuit. Arterial line pressures were also noted not to be higher than expected. The decision was taken not to replace the oxygenator and continue with CPB as there was no clinical indication for changing the oxygenator.

ACT's &  blood gases were repeated frequently throughout the case with no signs of  impairment of function. Towards the end of the bypass run the turbulent flow appeared to resolve & the RPM's also decreased. The oxygenator was  flushed through at the end of the case & there was significant residue  biological in nature in the filter.  

A total of 11 oxygenators were observed to have this occurrence. It is standard practice to shine a light directly on the oxygenator during CPB and to flush through the circuit to a cell saver.

No patients had a history of Covid 19 infection. In 2 of the cases coagulation screens were sent to our in hospital haematology lab, specifically D Dimer, Fibrinogen, APTT, PT, INR & PLT’s but no unusual results were noted.

DETAILS OF INJURY TO PATIENT

None

TYPE OF DEVICE/MANUFACTURER

Liva Nova Inspire 6F

ACTION TAKEN

The oxygenator was retained after the case & a full incident report was sent to Liva Nova. Pictures were taken during the case & following the flush through. 

This had happened previously on several occasions with no negative outcomes to the patient & were all reported to Liva Nova. A total of 3 oxygenators were returned for inspection. The manufacturer did not require further oxygenator returns.

Departmental decision was taken to send blood samples for a full coagulation screening to the hospital laboratory if it should occur again.

Manufacturer informed of the events and the oxygenator was retained for inspection.

Agreed to use 8F oxygenators going forward (update: since switching the same phenomenon has been experienced four times).

RESPONSE FROM MANUFACTURER

See report attached

ADDTIONAL INFORMATION

All cases where this was experienced were CABG's and only a single patient had confirmed COVID positive status at the time of procedure.

SUPPORTING VISUALS/VIDEO
SUPPORTING DOCUMENT

OTHER PERFUSION
 REPORTS

London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Select and scroll to view chosen report
REPORT NO
123
5/7/24
Hardware Fault
REPORT NO
122
5/4/24
Oxygenator Failure
REPORT NO
121
8/3/24
Disposables: Eurosets Adult ECMO Oxygenator
REPORT NO
120
1/5/24
Disposable Incident
REPORT NO
119
Hardware: HCU Contamination
REPORT NO
118
12/9/23
Human Factors: ECMO Mobilisation
REPORT NO
117
27/9/23
Disposables: Oxygenator
REPORT NO
116
14/9/23
Disposable Issue: Reservoir in CPB circuit
REPORT NO
115
13/9/23
Disposable Error
REPORT NO
114
13/7/23
Dispoables
REPORT NO
112
20/2/23
HLM Hardware Error: S5 HLM - Dual roller cardioplegia
REPORT NO
111
10/2/23
Human Factors
REPORT NO
110
31/1/23
DISPOSABLE FAULT
REPORT NO
109
19/1/23
Disposable issue
REPORT NO
108
3/1/23
Disposable issue: Oxygenator leak
REPORT NO
107
6/12/22
Hardware fault. Electronic gas Blender failure
REPORT NO
106
1/11/22
ECG Interference
REPORT NO
105
9/9/22
Issue with use of disposable
REPORT NO
104
13/6/22
Human Factors
REPORT NO
103
25/7/22
Human Factors
REPORT NO
102
7/7/22
Disposable issue
REPORT NO
101
6/4/21
Irregular blood flow/foreign looking substance in Oxygenator
REPORT NO
100
20/4/22
Gas Line Filter Fault
REPORT NO
099
20/4/22
Human Factors: BiVAD Error
REPORT NO
098
4/4/22
REPORT NO
097
19/3/22
Disposables Issue
REPORT NO
096
18/3/22
Bubbles Sensor Activation
REPORT NO
095
1/1/22
Protamine Administration
REPORT NO
094
1/11/21
Oxygenation failure upon commencement of CPB
REPORT NO
093
1/11/21
Detached bullet tip of sump-updated response Sep 2022
Perfusion Report cards are shown in batches of 30 at any one time where available.
123
122
121
120
119
118
117
116
115
114
112
111
110
109
108
107
106
105
104
103
102
101
100
099
098
097
096
095
094
093
092
091
090
089
088
087
086
085
084
083
082
081
080
079
078
077
076
075
074
073
072
071
070
069
068
067
066
065
064
063
062
061
060
059
058
057
056
055
054
053
052
051
050
049
048
047
046
045
044
043
042
041
040
039
038
037
036
035
034
033
032
031
030
029
028
027
026
025
024
023
London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Know the report number? Scroll horizontally through the circle icons and select to view the appropriate report.