As the Getinge HCU40 heater cooler was being cleaned and started defrosting, a popping sound, sparks and the smell of burning coming from the base of the heater cooler. The power was turned off immediately, estates and the company contacted, Datix completed.
n/a
Getinge HCU40
The vacuum valves were recently upgraded during service. Company will come reinstall the old valves (within a week) as the current vacuum valves are leaking onto the power board. This is a temporary measure until a better solution.
n/a
Please see FSN attached issued by Getinge regarding this issue. It is FSN 034 on the SCPS website.