PERFUSION REPORTS

REPORT
109
DATE
19/1/2023
TITLE
Disposable issue
DETAILS OF INCIDENT

While priming the Eurosets Adult A.L. one ECMO oxygenator (EU5019) it was noted that there was a leak from the bottom of the oxygenator. the leak appeared to originate form the seam between the fibres and the plastic housing, suspected lack of bonding. The Oxygenator was changed out during priming. A representative from the company was contacted and an inspection completed.

Eurosets testing:

We set-up the circuit: the reservoir is filled with 500 ml of physiological water; the oxygenator module is filled form the reservoir using a peristaltic pump (Q = 6 l/min) and remained in loop for 10 minutes. We notice a slow dripping in the lower part of the oxygenator, in front of the arterial blood out. To identify the exact point of leakage, the device was sectioned removing the Gas-escape lid, refilled with water, and then pressurized.

The point of loss occurred due to the fiber breakage of the last winding at the blood outlet.

The leak reported during the priming phase can be related to a fiber rupture occurred after Eurosets manufacturing and test phases. however, further improvement processes have been implemented to reduce this occurrence.

The safety committee have reviewed this incident and the safety report. it is acknowledged that the stringent checking by the perfusion team ensured that the leak was noted and the risk was appropriately managed. the Eurosets report outlines that they are in adherence to the appropriate controls and have met their risk matrix. However, the leak does question quality and a recognition by the community that disposable issues need to be assessed with adequate check.

DETAILS OF INJURY TO PATIENT

No injury

TYPE OF DEVICE/MANUFACTURER

Eurosets Adult A.L. one ECMO oxygenator (EU5019)

ACTION TAKEN

The device was removed and retained an sent off to the manufacturer for inspection.

RESPONSE FROM MANUFACTURER

Full report completed and actions taken.

Eurosets testing:

We set-up the circuit: the reservoir is filled with 500 ml of physiological water; the oxygenator module is filled form the reservoir using a peristaltic pump (Q = 6 l/min) and remained in loop for 10 minutes. We notice a slow dripping in the lower part of the oxygenator, in front of the arterial blood out. To identify the exact point of leakage, the device was sectioned removing the Gas-escape lid, refilled with water, and then pressurized.

The point of loss occurred due to the fiber breakage of the last winding at the blood outlet.

The leak reported during the priming phase can be related to a fiber rupture occurred after Eurosets manufacturing and test phases. however, further improvement processes have been implemented to reduce this occurrence

ADDTIONAL INFORMATION
SUPPORTING VISUALS/VIDEO
SUPPORTING DOCUMENT

OTHER PERFUSION
 REPORTS

London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Select and scroll to view chosen report
REPORT NO
123
5/7/24
Hardware Fault
REPORT NO
122
5/4/24
Oxygenator Failure
REPORT NO
121
8/3/24
Disposables: Eurosets Adult ECMO Oxygenator
REPORT NO
120
1/5/24
Disposable Incident
REPORT NO
119
Hardware: HCU Contamination
REPORT NO
118
12/9/23
Human Factors: ECMO Mobilisation
REPORT NO
117
27/9/23
Disposables: Oxygenator
REPORT NO
116
14/9/23
Disposable Issue: Reservoir in CPB circuit
REPORT NO
115
13/9/23
Disposable Error
REPORT NO
114
13/7/23
Dispoables
REPORT NO
112
20/2/23
HLM Hardware Error: S5 HLM - Dual roller cardioplegia
REPORT NO
111
10/2/23
Human Factors
REPORT NO
110
31/1/23
DISPOSABLE FAULT
REPORT NO
109
19/1/23
Disposable issue
REPORT NO
108
3/1/23
Disposable issue: Oxygenator leak
REPORT NO
107
6/12/22
Hardware fault. Electronic gas Blender failure
REPORT NO
106
1/11/22
ECG Interference
REPORT NO
105
9/9/22
Issue with use of disposable
REPORT NO
104
13/6/22
Human Factors
REPORT NO
103
25/7/22
Human Factors
REPORT NO
102
7/7/22
Disposable issue
REPORT NO
101
6/4/21
Irregular blood flow/foreign looking substance in Oxygenator
REPORT NO
100
20/4/22
Gas Line Filter Fault
REPORT NO
099
20/4/22
Human Factors: BiVAD Error
REPORT NO
098
4/4/22
REPORT NO
097
19/3/22
Disposables Issue
REPORT NO
096
18/3/22
Bubbles Sensor Activation
REPORT NO
095
1/1/22
Protamine Administration
REPORT NO
094
1/11/21
Oxygenation failure upon commencement of CPB
REPORT NO
093
1/11/21
Detached bullet tip of sump-updated response Sep 2022
Perfusion Report cards are shown in batches of 30 at any one time where available.
123
122
121
120
119
118
117
116
115
114
112
111
110
109
108
107
106
105
104
103
102
101
100
099
098
097
096
095
094
093
092
091
090
089
088
087
086
085
084
083
082
081
080
079
078
077
076
075
074
073
072
071
070
069
068
067
066
065
064
063
062
061
060
059
058
057
056
055
054
053
052
051
050
049
048
047
046
045
044
043
042
041
040
039
038
037
036
035
034
033
032
031
030
029
028
027
026
025
024
023
London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Know the report number? Scroll horizontally through the circle icons and select to view the appropriate report.