PERFUSION REPORTS

REPORT
052
DATE
1/3/2019
TITLE
Gas Line Filter Issue
DETAILS OF INCIDENT

On-going problem with Liva Nova green gas line with filter since November 2018 (at least 10 gas lines that reported department are aware of).  

Noticed pre-CPB when checking gas flow through gas line, pressure readings over 30mmHg on Maquet electronic gas blender.  When gas sweep increased it alarmed at 50mmHg.  When the gas line filter was disconnected the pressure returned to 0.

This has happened numerous times when using different Maquet electronic gas blenders (EGB 40).  It was not noted when using Sechrist blenders as pressure readings are not available on these blenders.  

DETAILS OF INJURY TO PATIENT

None

TYPE OF DEVICE/MANUFACTURER

Liva Nova gas line

Product Code: CP8049

Ref: 08049

ACTION TAKEN

Reported to Liva Nova. 6 gas lines that were deemed faulty were collected by Liva Nova for further investigation.

Liva Nova have supplied an alternative gas line filter (Pall) in case further issues arise while using the Maquet electronic gasblender.  When using the Sechrist blenders and the Liva Nova custom pack the gas line is automatically swapped for a Pall gas line.  

RESPONSE FROM MANUFACTURER

See attached below

ADDTIONAL INFORMATION

Upon reflection what do you think went well?

The issues was always identified prior to going on CPB so no patient was put at risk.  Members of the perfusion team were informed quickly so all were aware and able to keep defective items.  It was reported quickly to Liva Nova.  

The team reflected on the use of gas blenders without the ability to notice pressure changes within the line.  

New action plans/changes as a result of the event?

Liva Nova custom packs going forward will only be supplied with Pall gas line filter.  

SUPPORTING VISUALS/VIDEO
SUPPORTING DOCUMENT

OTHER PERFUSION
 REPORTS

London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Select and scroll to view chosen report
REPORT NO
123
5/7/24
Hardware Fault
REPORT NO
122
5/4/24
Oxygenator Failure
REPORT NO
121
8/3/24
Disposables: Eurosets Adult ECMO Oxygenator
REPORT NO
120
1/5/24
Disposable Incident
REPORT NO
119
Hardware: HCU Contamination
REPORT NO
118
12/9/23
Human Factors: ECMO Mobilisation
REPORT NO
117
27/9/23
Disposables: Oxygenator
REPORT NO
116
14/9/23
Disposable Issue: Reservoir in CPB circuit
REPORT NO
115
13/9/23
Disposable Error
REPORT NO
114
13/7/23
Dispoables
REPORT NO
112
20/2/23
HLM Hardware Error: S5 HLM - Dual roller cardioplegia
REPORT NO
111
10/2/23
Human Factors
REPORT NO
110
31/1/23
DISPOSABLE FAULT
REPORT NO
109
19/1/23
Disposable issue
REPORT NO
108
3/1/23
Disposable issue: Oxygenator leak
REPORT NO
107
6/12/22
Hardware fault. Electronic gas Blender failure
REPORT NO
106
1/11/22
ECG Interference
REPORT NO
105
9/9/22
Issue with use of disposable
REPORT NO
104
13/6/22
Human Factors
REPORT NO
103
25/7/22
Human Factors
REPORT NO
102
7/7/22
Disposable issue
REPORT NO
101
6/4/21
Irregular blood flow/foreign looking substance in Oxygenator
REPORT NO
100
20/4/22
Gas Line Filter Fault
REPORT NO
099
20/4/22
Human Factors: BiVAD Error
REPORT NO
098
4/4/22
REPORT NO
097
19/3/22
Disposables Issue
REPORT NO
096
18/3/22
Bubbles Sensor Activation
REPORT NO
095
1/1/22
Protamine Administration
REPORT NO
094
1/11/21
Oxygenation failure upon commencement of CPB
REPORT NO
093
1/11/21
Detached bullet tip of sump-updated response Sep 2022
Perfusion Report cards are shown in batches of 30 at any one time where available.
123
122
121
120
119
118
117
116
115
114
112
111
110
109
108
107
106
105
104
103
102
101
100
099
098
097
096
095
094
093
092
091
090
089
088
087
086
085
084
083
082
081
080
079
078
077
076
075
074
073
072
071
070
069
068
067
066
065
064
063
062
061
060
059
058
057
056
055
054
053
052
051
050
049
048
047
046
045
044
043
042
041
040
039
038
037
036
035
034
033
032
031
030
029
028
027
026
025
024
023
London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Know the report number? Scroll horizontally through the circle icons and select to view the appropriate report.