PERFUSION REPORTS

REPORT
018
DATE
TITLE
Medos Heat Exchanger Leak
DETAILS OF INCIDENT

During cardiopulmonary bypass (MVR+CABG) blood was seen in the water lines that connect the heater cooler unit to the oxygenator’s heat exchanger. This obviously suggests a leak in the heat exchanger and the patients’ blood and water could be potentially mixing. The water lines were immediately switched off and disconnected, and the consultant surgeon informed.
The recommended procedure is to change out the oxygenator following heat exchanger leak. However, the consultant cardiac surgeon felt that this was not a safe option at the time as the oxygenator was still working, so we remained on bypass. The cross clamp was removed and the patient was rewarmed from 28°C to 34°C using the separate heat exchanger on the cardioplegia circuit and two fluid warmers donated by the anaesthetic team. Once the patient reached 34°C, the patient was ventilated and bypass was terminated and the oxygenator replaced with a new one. Bypass was re-commenced and the patient fully rewarmed. Bypass was discontinued and the patient was stable. It should be noted that a full heat exchanger leak test was performed before the bypass circuit was primed and no fault was detected (water circulated at 37oC for 5mins prior to adding prime solution, no pressure test).

DETAILS OF INJURY TO PATIENT

Unknown. Patient stable and extubated the following morning.

TYPE OF DEVICE/MANUFACTURER

N/A

ACTION TAKEN

Microbiology was informed and antibiotics were given. A water sample from the heater cooler unit was sent to the lab for investigation, and the oxygenator was sent to Medos. A Datix report has been completed and MHRA informed by Medos.

RESPONSE FROM MANUFACTURER

N/A

ADDTIONAL INFORMATION

N/A

SUPPORTING VISUALS/VIDEO

N/A

SUPPORTING DOCUMENT

OTHER PERFUSION
 REPORTS

London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Select and scroll to view chosen report
REPORT NO
002
1/1/10
Luer Cap Found In Venous Inlet Port
REPORT NO
001
1/1/09
Oxygenator Water To Blood Leak
Perfusion Report cards are shown in batches of 30 at any one time where available.
123
122
121
120
119
118
117
116
115
114
112
111
110
109
108
107
106
105
104
103
102
101
100
099
098
097
096
095
094
093
092
091
090
089
088
087
086
085
084
083
082
081
080
079
078
077
076
075
074
073
072
071
070
069
068
067
066
065
064
063
062
061
060
059
058
057
056
055
054
053
052
051
050
049
048
047
046
045
044
043
042
041
040
039
038
037
036
035
034
033
032
031
030
029
028
027
026
025
024
023
London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Know the report number? Scroll horizontally through the circle icons and select to view the appropriate report.