Chalice Paragon Heat Exchanger Water To Blood Leak
DETAILS OF INCIDENT

During cardiopulmonary bypass (AVR) the perfusionist noticed blood in the water lines that connect the heater cooler unit to the oxygenator's (paragon) heat exchanger 37 minutes in to bypass. This suggests a leak in the heat exchanger and the patients' blood and water could be mixing.

The water lines were immediately switched off and disconnected, and the consultant surgeon informed. The patient remained on bypass while a new bypass circuit was prepared. Bypass was temporarily stopped and re-commenced with the new circuit using a new heater-cooler and the patient fully rewarmed. Bypass was discontinued and the patient was stable. It should be noted that a full heat exchanger leak test was performed before the bypass circuit was primed and no fault was detected (water circulated at 37oC for 5mins prior to adding prime solution, no pressure test). Blood and water samples were sent for testing. Antibiotics were given by the consultant anaesthetist.

DETAILS OF INJURY TO PATIENT

Patient stable and extubated following day. Blood samples sterile to date.

TYPE OF DEVICE/MANUFACTURER

N/A

ACTION TAKEN

Microbiology was informed and antibiotics were given. A water sample from the heater cooler unit was sent to the lab for investigation, and the oxygenator was sent to Chalice.

RESPONSE FROM MANUFACTURER

A Datix report has been completed and MHRA informed by Chalice.

ADDTIONAL INFORMATION

N/A

SUPPORTING VISUALS/VIDEO

N/A

SUPPORTING DOCUMENT

OTHER PERFUSION
 REPORTS

London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Select and scroll to view chosen report
REPORT NO
002
1/1/10
Luer Cap Found In Venous Inlet Port
REPORT NO
001
1/1/09
Oxygenator Water To Blood Leak
Perfusion Report cards are shown in batches of 30 at any one time where available.
123
122
121
120
119
118
117
116
115
114
112
111
110
109
108
107
106
105
104
103
102
101
100
099
098
097
096
095
094
093
092
091
090
089
088
087
086
085
084
083
082
081
080
079
078
077
076
075
074
073
072
071
070
069
068
067
066
065
064
063
062
061
060
059
058
057
056
055
054
053
052
051
050
049
048
047
046
045
044
043
042
041
040
039
038
037
036
035
034
033
032
031
030
029
028
027
026
025
024
023
London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Know the report number? Scroll horizontally through the circle icons and select to view the appropriate report.