Paragon Oxygenator Gas Inlet Leak
DETAILS OF INCIDENT

After 2 days on ECMO, the paragon oxygenator started to leak gas from around the oxygenator gas inlet top plate.

DETAILS OF INJURY TO PATIENT

This showed as reduced patient SaO2 and poor post oxygenator blood gases.

TYPE OF DEVICE/MANUFACTURER

N/A

ACTION TAKEN

This is the second such leak in 2 weeks, same batch number. All other products with this batch number has been removed from stock and quarantined for collection by manufacturer.

RESPONSE FROM MANUFACTURER

N/A

ADDTIONAL INFORMATION

N/A

SUPPORTING VISUALS/VIDEO

N/A

SUPPORTING DOCUMENT

OTHER PERFUSION
 REPORTS

London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Select and scroll to view chosen report
REPORT NO
003
1/2/10
Difficulties In Visualising Blender Reading
REPORT NO
002
1/1/10
Luer Cap Found In Venous Inlet Port
REPORT NO
001
1/1/09
Oxygenator Water To Blood Leak
Perfusion Report cards are shown in batches of 30 at any one time where available.
123
123
122
121
120
119
118
117
116
115
114
112
111
110
109
108
107
106
105
104
103
102
101
100
099
098
097
096
095
094
093
092
091
090
089
088
087
086
085
084
083
082
081
080
079
078
077
076
075
074
073
072
071
070
069
068
067
066
065
064
063
062
061
060
059
058
057
056
055
054
053
052
051
050
049
048
047
046
045
044
043
042
041
040
039
038
037
036
035
034
033
032
031
030
029
028
027
026
025
024
London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Know the report number? Scroll horizontally through the circle icons and select to view the appropriate report.