Faulty Connector On ECMO Soft Shell Reservoir Bag
DETAILS OF INCIDENT

The units custom-made ECMO circuit contains a soft shell reservoir bag to assist with priming and de-airing. During the set-up of an ECMO circuit crystalloid fluid was added to the reservoir bag and the centrifugal head, oxygenator and associated tubing circuit primed passively. After correct positioning of the head in the system driver and applying power to the circuit no flow could be produced. The circuit was checked for obstruction and on close inspection it was found that the outlet of the bag was almost completely adhered to the bag structure. Cavitation was observed in the centrifugal head inlet line. The application of negative pressure to the outlet effectively sealed the bag preventing flow. The seal was pulled apart sufficiently enough to allow flow however slight damage to the bag occurred. The bag is always removed from the circuit prior to the initiation of ECMO and after adequate de-airing this was performed in the usual manner. ECMO was then established in the patient uneventfully.

DETAILS OF INJURY TO PATIENT

None

TYPE OF DEVICE/MANUFACTURER

N/A

ACTION TAKEN

The bag was kept and returned to the manufacturer.
 The other members of the department were made aware of the incident so that the appropriate checks could be made to subsequent ECMO circuits when used.


RESPONSE FROM MANUFACTURER

The following response was received:

"This is an isolated incident, unfortunately caused by an operator error during manufacture. The connector was over bonded when inserted into the outlet at the bottom of the reservoir bag where the bonding together occurred. Operatives have been made fully aware of the issue, an extra quality check along with details of this complaint have been added to the relevant stage of manufacture, which should prevent reoccurrence"



ADDTIONAL INFORMATION

N/A

SUPPORTING VISUALS/VIDEO

N/A

SUPPORTING DOCUMENT

OTHER PERFUSION
 REPORTS

London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Select and scroll to view chosen report
REPORT NO
002
1/1/10
Luer Cap Found In Venous Inlet Port
REPORT NO
001
1/1/09
Oxygenator Water To Blood Leak
Perfusion Report cards are shown in batches of 30 at any one time where available.
123
122
121
120
119
118
117
116
115
114
112
111
110
109
108
107
106
105
104
103
102
101
100
099
098
097
096
095
094
093
092
091
090
089
088
087
086
085
084
083
082
081
080
079
078
077
076
075
074
073
072
071
070
069
068
067
066
065
064
063
062
061
060
059
058
057
056
055
054
053
052
051
050
049
048
047
046
045
044
043
042
041
040
039
038
037
036
035
034
033
032
031
030
029
028
027
026
025
024
023
London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Know the report number? Scroll horizontally through the circle icons and select to view the appropriate report.