PERFUSION REPORTS

REPORT
034
DATE
TITLE
Possible Medtronic Fusion Oxygenator Failure
DETAILS OF INCIDENT

60 year old 80 kg male patient. Blood group B+ve. Previous PCI During coronary artery bypass grafting surgery, once Cardiopulmonary bypass (CPB) commenced, there were very high pre-oxygenator P1 line pressures in the bypass circuit (450mmHg). The oxygenator was a Medtronic Fusion. P2 post-oxygenator pressures were within 'normal' range. All potential causes were eliminated but it persised.

The high P1 pressures were not resolving and the p02 oxygen measurement started to fall gradually, although not compromising patient oxygenation. Clotting levels were within range and no clots were seen in the oxygenator. The surgeon was kept informed and the situation monitored before progressing with operation. The decision was made to change out the oxygenator. The anaesthetist was informed and the patient was ventilated conventionally.

CPB was stopped and the oxygenator changed. CPB recommenced and the same high P1 pressures seen again but with no compromise to oxygenation so bypass was continued and monitored very closely. After 30 minutes, the pressures fell to 'normal' levels and remained that way for remainder of the bypass. It was also noted that patient haematocrit on CPB was high (38%) and following discussion with the anaesthetist the patient was diluted with the addition of crystalloid.

Post CPB and protamine, a distal vein graft needed repair and heparin was administered and CPB recommenced. P1 and P2 were both within 'normal levels'. As high P1 pressures seen with both devices on initial use, this was suggestive of a patient specific incident. No untoward effects were noted in the post-op period.

DETAILS OF INJURY TO PATIENT

N/A

TYPE OF DEVICE/MANUFACTURER

N/A

ACTION TAKEN

First oxygenator returned to Medtronic for evaluation. No fault was found with the oxygenator.

In-house discussion:

Following the incident there was a feeling that perhaps the oxygenator was changed out too soon and high pre oxygenator pressures would have stabilised.


RESPONSE FROM MANUFACTURER

N/A

ADDTIONAL INFORMATION

N/A

SUPPORTING VISUALS/VIDEO

N/A

SUPPORTING DOCUMENT

OTHER PERFUSION
 REPORTS

London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Select and scroll to view chosen report
REPORT NO
123
5/7/24
Hardware Fault
REPORT NO
122
5/4/24
Oxygenator Failure
REPORT NO
121
8/3/24
Disposables: Eurosets Adult ECMO Oxygenator
REPORT NO
120
1/5/24
Disposable Incident
REPORT NO
119
Hardware: HCU Contamination
REPORT NO
118
12/9/23
Human Factors: ECMO Mobilisation
REPORT NO
117
27/9/23
Disposables: Oxygenator
REPORT NO
116
14/9/23
Disposable Issue: Reservoir in CPB circuit
REPORT NO
115
13/9/23
Disposable Error
REPORT NO
114
13/7/23
Dispoables
REPORT NO
112
20/2/23
HLM Hardware Error: S5 HLM - Dual roller cardioplegia
REPORT NO
111
10/2/23
Human Factors
REPORT NO
110
31/1/23
DISPOSABLE FAULT
REPORT NO
109
19/1/23
Disposable issue
REPORT NO
108
3/1/23
Disposable issue: Oxygenator leak
REPORT NO
107
6/12/22
Hardware fault. Electronic gas Blender failure
REPORT NO
106
1/11/22
ECG Interference
REPORT NO
105
9/9/22
Issue with use of disposable
REPORT NO
104
13/6/22
Human Factors
REPORT NO
103
25/7/22
Human Factors
REPORT NO
102
7/7/22
Disposable issue
REPORT NO
101
6/4/21
Irregular blood flow/foreign looking substance in Oxygenator
REPORT NO
100
20/4/22
Gas Line Filter Fault
REPORT NO
099
20/4/22
Human Factors: BiVAD Error
REPORT NO
098
4/4/22
REPORT NO
097
19/3/22
Disposables Issue
REPORT NO
096
18/3/22
Bubbles Sensor Activation
REPORT NO
095
1/1/22
Protamine Administration
REPORT NO
094
1/11/21
Oxygenation failure upon commencement of CPB
REPORT NO
093
1/11/21
Detached bullet tip of sump-updated response Sep 2022
Perfusion Report cards are shown in batches of 30 at any one time where available.
123
122
121
120
119
118
117
116
115
114
112
111
110
109
108
107
106
105
104
103
102
101
100
099
098
097
096
095
094
093
092
091
090
089
088
087
086
085
084
083
082
081
080
079
078
077
076
075
074
073
072
071
070
069
068
067
066
065
064
063
062
061
060
059
058
057
056
055
054
053
052
051
050
049
048
047
046
045
044
043
042
041
040
039
038
037
036
035
034
033
032
031
030
029
028
027
026
025
024
023
London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Know the report number? Scroll horizontally through the circle icons and select to view the appropriate report.