PERFUSION REPORTS

REPORT
040
DATE
1/10/2018
TITLE
Reservoir Clot
DETAILS OF INCIDENT

The patient was 15.3kg and presented requiring re-sternotomy for LVOTO relief. An appropriate size oxygenator and tubing combination were selected, and the circuit primed with a blood, albumin, Hartmanns mix and 2500in heparin. Following a test-dose aprotinin was added to the circuit.

Prior to the initiation of cardiopulmonary bypass, the baseline ACT was within normal range. Following a heparin dose of 400iu/kg (6,000iu) the ACT was sub optimal for bypass at 367 seconds. After discussion with the anaesthetist a further 1,000 units of heparin was administered to the patient and ACT repeated. Subsequent ACT was again sub-optimal at 409 seconds. Further discussion took place with the anaesthetist and the two perfusionists for the case regarding the need for additional heparin doses. It was agreed that this was an adequate level of anticoagulation to proceed with bypass. The unit protocol states an ACT of 480 seconds is used for CPB.

On commencement of bypass the first ACT was below the required level at 254 seconds. A further 2,000 units of heparin was administered to the patient. ACT was repeated after adequate time to circulate through the system and was again low at 458 seconds. Therefore, a further dose of 1,500 units of heparin was administered. Repeated ACT showed this to be within the normal range for bypass, in excess of 500 seconds. At this point it was noted that venous levels were dropping dramatically, and it appeared that the venous filter was clotting off. A discussion took place between the perfusionist and the surgeon as to the course of action to be taken. It was decided the venous reservoir should be changed out before proceeding any further with the operation. Patient was prepared for coming off bypass by rewarming and filling the heart to achieve adequate cardiac output. The venous reservoir was changed out uneventfully, and cardiopulmonary bypass recommenced without any issues. Subsequent ACTs during the case were all in excess of 500 seconds. The operation was completed, and the patient weaned successfully. The patient recovered fully, and no untoward effects were noted.

Note: When the case was performed the unit had recently transitioned from using Haemochron celite tubes to Haemochron junior cartridges for ACT determination. This change in technique is widely known (and reported in the analyser information) to produce lower measurements for a sample of blood. This information was taken into account during the decision making.

DETAILS OF INJURY TO PATIENT

None

TYPE OF DEVICE/MANUFACTURER

N/A

ACTION TAKEN

Learning Outcomes:

Following the case discussions were made regarding whether the patient was inadequately heparinised or whether another factor has caused the reservoir clot to develop e.g. suction, unknown patient factor, anti-thrombin III deficiency.

The case did deviate from the protocol, but it was felt appropriate in the context of the situation. This has stimulated a departmental review of practice particularly in view of introducing a new, different measuring technique and whether the protocol should be altered because of this.

RESPONSE FROM MANUFACTURER

N/A

ADDTIONAL INFORMATION

N/A

SUPPORTING VISUALS/VIDEO

N/A

SUPPORTING DOCUMENT

OTHER PERFUSION
 REPORTS

London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Select and scroll to view chosen report
REPORT NO
062
1/8/20
Human Factors Issue
REPORT NO
061
1/8/20
LIVA NOVA RESPONSE TO REPORTED CENTRIFUGAL HEAD ISSUE
REPORT NO
060
1/8/20
Oxygenator Fibre Leak
REPORT NO
059
1/5/20
Failure to remove CO2 effectively
REPORT NO
058
1/7/20
Centrifugal Head Issue 3
REPORT NO
057
1/6/20
Centrifugal Head Issue 2
REPORT NO
056
1/1/20
Centrifugal Head Issue
REPORT NO
055
1/2/20
Faulty Pressure Relief Valves
REPORT NO
054
1/4/20
Leaking ALF
REPORT NO
053
1/2/20
Arterial Line Filter Leak
REPORT NO
052
1/3/19
Gas Line Filter Issue
REPORT NO
051
1/12/19
Dual Lumen Cannula Movement
REPORT NO
050
1/10/19
Venous Cannula Damaged
REPORT NO
049
1/12/19
Oxygenation Issue
REPORT NO
048
1/8/19
Roller Pump Failure
REPORT NO
047
1/12/19
Broken Cannula Dilator
REPORT NO
046
6/12/19
Gas Blender Error
REPORT NO
045
20/11/19
Tubing Detachments
REPORT NO
044
27/11/19
Interruption to IABP counterpulsation
REPORT NO
043
1/7/19
Getinge Mal-Position Of IABP Catheter
REPORT NO
042
1/2/19
Avalon Dual Lumen Cannula Crack
REPORT NO
041
1/2/19
Custodial Rubber Fragmentation In Spiked Bag
REPORT NO
040
1/10/18
Reservoir Clot
REPORT NO
039
1/10/18
Pall Clot In Cardioplegia Circuit
REPORT NO
038
1/10/18
Maquet HL20 Cardioplegia Pump Error
REPORT NO
037
1/10/17
Temed D200 Haemoconcentrator Clot
REPORT NO
036
1/10/17
Oxygenator issue
REPORT NO
035
1/10/17
Medtronic Fusion Oxygenator Blood To Gas Leak
REPORT NO
034
Possible Medtronic Fusion Oxygenator Failure
REPORT NO
033
Fannin Heparin Ampules Shattering
Perfusion Report cards are shown in batches of 30 at any one time where available.
123
122
121
120
119
118
117
116
115
114
112
111
110
109
108
107
106
105
104
103
102
101
100
099
098
097
096
095
094
093
092
091
090
089
088
087
086
085
084
083
082
081
080
079
078
077
076
075
074
073
072
071
070
069
068
067
066
065
064
063
062
061
060
059
058
057
056
055
054
053
052
051
050
049
048
047
046
045
044
043
042
041
040
039
038
037
036
035
034
033
032
031
030
029
028
027
026
025
024
023
London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Know the report number? Scroll horizontally through the circle icons and select to view the appropriate report.