PERFUSION REPORTS

REPORT
049
DATE
1/12/2019
TITLE
Oxygenation Issue
DETAILS OF INCIDENT

Background

Patient with aortic stenosis and coronary artery disease for AVR+CABGx2(LIMA+SVG).  

Patient details: Height 179 cms, Weight 111 kg, Cardiac Output 5.5LPM with cardiac index of 2.4.

Pre-op Hb 14.4g/dl

Prime Volume 800mls Hartmanns, 5000IU Heparin, 4mmol Magnesium

APP/RAP carried out, target temp 32 degrees Celsius.  

CO2 flushing at 5LPM was being utilised, this was reduced upon pCO2 increase to help attenuate to no avail.  

Details of Incident

The first 60 mins of bypass were uneventful with average pO2’s of 25kPa.  Approximately 60 mins into CPB patient pO2 and pCO2 began to fall/rise respectively.  Sweep gas and FIO2 were increased to compensate.  However both pO2 and pCO2 were out of protocol range despite running a sweep of 9LPM and FIO2 of100%.  

A 20L gas blender was inserted into the circuit to compensate, a sweep of 15 LPM was required to attenuate pCO2.

pO2 remained below protocol minimums however they were within physiological normal range (for the whole case, the lowest pO2 recorded was 11.35kPa).  

Venous saturation dropped to below 60% and lactate slowly increased.  Pump flows averaged 100% throughout.  

Hb averaged 12.65g/dl, DiO2 387m/min/m2 and ACT’s were above 400 seconds and any drops were treated with 5000 IU heparin (lowest ACT on bypass 379 seconds).  Transmembrane pressures were not measured as the unit routinely uses centrifugal pumps. The RPM flow rate did not decrease, nor was there any indication of clot formation upon inspection of oxygenator post CPB.  

The issues being experienced were communicated with the surgeon and the potential need for an oxygenator changeout was discussed.  The surgeon was close to removing the cross clamp and the decision was made to wait.  Once the cross clamp was removed, ventilation was resumed and the heart was allowed to eject. Blood gases and venous saturation began to slowly improve until CPB was terminated

DETAILS OF INJURY TO PATIENT

None

TYPE OF DEVICE/MANUFACTURER

Liva Nova 6F Oxygenator

ACTION TAKEN

20L gas blender was inserted to attenuate high pCO2.  Replacement oxygenator was statically primed should there have been a further drop in oxygenation status, thus prompting a change out.  Ventilation was resumed after removing of the cross clamp and the heart was allowed to eject for the duration until CPB had been terminated.  

Consideration was given to a gas filter failure and the filter was removed to no avail.  

The oxygenator was retained and sent back to Liva Nova for investigation.  

RESPONSE FROM MANUFACTURER

n/a

ADDTIONAL INFORMATION

Post-operatively, there was some speculation within the department that it may not have been the oxygenator but rather a hypermetabolic state, specifically malignant hyperthermia (MH), however the only indication on CPB was hypercarbia requiring a large amount of sweep to correct and falling venous saturation despite a physiologically normal pO2 as per Butala et al.2018.  

Patient temperature did not rise acutely but this would have been confounded with the use of a heater cooler.  The patient did not present with symptoms of MH on induction of anaesthesia and no dantrolene was administered.  

Post event the patient was sent for MH testing-awaiting update on results.  

Upon reflection what do you think went well?

The decision not to change out the oxygenator and to wait until the cross clamp came off to allow the heart to eject.  

SUPPORTING VISUALS/VIDEO

n/a

SUPPORTING DOCUMENT

OTHER PERFUSION
 REPORTS

London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Select and scroll to view chosen report
REPORT NO
092
1/10/21
Vapourizer Issue
REPORT NO
091
1/10/21
Oyxgenator Leak
REPORT NO
090
1/10/21
Breach of chemical cleaner
REPORT NO
089
1/9/21
Sechrist Gas Blender Sweep on ECMO
REPORT NO
088
1/9/21
Oxygenator High Pressure Events
REPORT NO
087
1/10/21
Leaking one way valve
REPORT NO
086
1/7/21
Electronic Gas Blender Fault
REPORT NO
085
1/2/21
Protamine Administration Issue
REPORT NO
084
1/6/21
Damaged roller pump lid
REPORT NO
083
1/6/21
IABP malfunction
REPORT NO
082
1/1/21
Accidental cardioplegia administration
REPORT NO
081
1/4/21
S5 Double Roller Pump incident
REPORT NO
080
1/5/21
Human factors issue
REPORT NO
079
1/6/21
Centrimag battery module failure
REPORT NO
078
1/6/21
Sampling manifold valve fault
REPORT NO
077
1/5/21
IABP transport issue
REPORT NO
076
1/4/21
Modified Ultrafiltration (MUF) Issue
REPORT NO
075
1/3/21
Electromagnetic interference with HLM
REPORT NO
074
1/2/21
HCU40 Failure
REPORT NO
073
1/2/21
HCU40 valve failures
REPORT NO
072
1/1/21
HCU40 failure
REPORT NO
071
1/1/21
LIVA NOVA STATEMENT ON CENTRIFUGAL HEAD
REPORT NO
070
1/1/21
Loose component on circuit
REPORT NO
069
1/12/20
Centrifugal Head noise
REPORT NO
068
1/11/20
Venous Reservoir disposable issue
REPORT NO
067
1/11/20
Failure of Heater Cooler due to leaking valves
REPORT NO
066
1/11/20
Failure of Valves on heater cooler unit
REPORT NO
065
1/9/20
Intra Aortic Balloon Pump (IABP) Failure
REPORT NO
064
1/9/20
Centrifugal head failure
REPORT NO
063
1/8/20
Centrifugal head issue 4
Perfusion Report cards are shown in batches of 30 at any one time where available.
123
122
121
120
119
118
117
116
115
114
112
111
110
109
108
107
106
105
104
103
102
101
100
099
098
097
096
095
094
093
092
091
090
089
088
087
086
085
084
083
082
081
080
079
078
077
076
075
074
073
072
071
070
069
068
067
066
065
064
063
062
061
060
059
058
057
056
055
054
053
052
051
050
049
048
047
046
045
044
043
042
041
040
039
038
037
036
035
034
033
032
031
030
029
028
027
026
025
024
023
London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Know the report number? Scroll horizontally through the circle icons and select to view the appropriate report.