PERFUSION REPORTS

REPORT
051
DATE
1/12/2019
TITLE
Dual Lumen Cannula Movement
DETAILS OF INCIDENT

A neonatal patient was placed on VV ECMO for acute hypoxic respiratory failure.  The patient was unstable and cannulated bedside on the PICU under echo guidance.  X-ray confirmed correct placement of the cannula.  The cannula was secured in the standard method for neck cannulation.  

After 3 days of therapy, it was noticed the recirculation increased significantly from 1-3% to 45%.  

During physiotherapy sessions, which included proning the patient, the cannula slipped out of the IVC and was free in the RA.  After discussions with an MDT, the balance of risk was discussed about recannulating.

There are case studies of repositioning the cannula with a “lasso” from the IVC but in this case it was decided to leave the cannula as the patient was improving and it was anticipated that they would be weaned the following day.  

The patient failed to wean from ECMO and was kept on ECMO.  1 day later the cannula pierced the RA causing a tamponade and required surgery and conversion to VA ECMO.  The patient was successfully weaned 3 days later.  

DETAILS OF INJURY TO PATIENT

None

TYPE OF DEVICE/MANUFACTURER

Maquet Avalon Elite Bi-Caval Dual Lumen Cannula

Product Code: 10016-CE

Lot/Batch: 246799

ACTION TAKEN

Reported to the company.

A discussion was had with a peer centre to ask if they would have had the same response. The peer centre has had similar translocations. They would observe in a similar situation. They did have a puncture in one of their migrated cannula too.

RESPONSE FROM MANUFACTURER

n/a

ADDTIONAL INFORMATION

Upon reflection whatdo you think went well?

The junior team provided excellent support whilst the senior on-call team came in.  This included a pericardiocentesis.  

New action plans/changes as a result of the event?

The team are very aware of the risks with the cannula for VV and so may have taken a more positive look at VA ECMO.  Although a more risky procedure for respiratory ECMO, the exposure to VA means that is managed with better experience than VV.  

SUPPORTING VISUALS/VIDEO
SUPPORTING DOCUMENT

OTHER PERFUSION
 REPORTS

London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Select and scroll to view chosen report
REPORT NO
032
Sucker Placed In Raceway In Wrong Direction
REPORT NO
031
High Transmembrane Pressure Gradients
REPORT NO
030
Faulty Connector On ECMO Soft Shell Reservoir Bag
REPORT NO
029
Sorin Inspire 8F Blood Leak From Temperature Port
REPORT NO
028
Sorin Inspire 8F Oxygenator Detached From Holding Arm
REPORT NO
027
Paragon Oxygenator Gas Inlet Leak
REPORT NO
026
Maquet IABP Console Failure
REPORT NO
025
Sorin EOS Gas To Blood Leak Whilst Priming
REPORT NO
024
Chalice Paragon Midi Heat Exchanger Leak Whilst Priming
REPORT NO
023
Chalice Paragon Heat Exchanger Water To Blood Leak
REPORT NO
022
Vent Pump Boot Tubing Inserted In Wrong Way Round
REPORT NO
021
Paragon Oxygenator Heat Exchanger Leak
REPORT NO
020
S3 E62 Pump Error
REPORT NO
019
Medtronic Potential Oxygenator Leak
REPORT NO
018
Medos Heat Exchanger Leak
REPORT NO
017
Avant D903 Has A Dual Chamber Reservoir Leak
REPORT NO
016
Additional Sucker Incorrectly Placed In Raceway
REPORT NO
015
Large Perished Tear Observed On Tubing At Back Of Vaporiser
REPORT NO
014
Gas Blender Flow Control Immobilised
REPORT NO
013
Isoflurane Vaporiser Seal Leak Due To Malposition
REPORT NO
012
1/2/13
Hypercoagulable Disorder May Have Led To Oxygenator Failure
REPORT NO
011
1/1/13
LV Vent Placed In Raceway In Wrong Direction
REPORT NO
010
1/3/12
Sucker Placed In Raceway In Wrong Direction
REPORT NO
009
1/2/12
Similarity In Heparin And St Thomas Cardioplegia Ampules
REPORT NO
008
1/7/11
Aorta Remained Clamped Post CPB
REPORT NO
007
1/5/11
Deficiencies In The QC At Terumo Ann Arbor, USA.
REPORT NO
006
1/4/11
Awareness Of Suckers When Using Bio-Glues
REPORT NO
005
1/2/11
Patient Exsanguinated From Additional Venous Line Post CPB
REPORT NO
004
1/1/11
Possible Inaccuracies In Colloid Research
REPORT NO
003
1/2/10
Difficulties In Visualising Blender Reading
Perfusion Report cards are shown in batches of 30 at any one time where available.
123
122
121
120
119
118
117
116
115
114
112
111
110
109
108
107
106
105
104
103
102
101
100
099
098
097
096
095
094
093
092
091
090
089
088
087
086
085
084
083
082
081
080
079
078
077
076
075
074
073
072
071
070
069
068
067
066
065
064
063
062
061
060
059
058
057
056
055
054
053
052
051
050
049
048
047
046
045
044
043
042
041
040
039
038
037
036
035
034
033
032
031
030
029
028
027
026
025
024
023
London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Know the report number? Scroll horizontally through the circle icons and select to view the appropriate report.