PERFUSION REPORTS

REPORT
078
DATE
1/6/2021
TITLE
Sampling manifold valve fault
DETAILS OF INCIDENT

On CPB, the flow through the manifold samples line was severely restricted. It was noticed during administration through the line as it was easy to see clear fluid flowing extremely slowly along the line.

Also, the temperature read by the CDI monitoring was affected, with readings of around 6degrees Celsius lower than the actual one, with effects on pH stat values.

By removing the one way valve, all issues were resolved promptly.

DETAILS OF INJURY TO PATIENT

None

TYPE OF DEVICE/MANUFACTURER

Dideco D101 Neonatal and Paediatric Set

ACTION TAKEN

All staff informed and warned to check for same issues. One way valve kept and packaged for return to manufacturer.

RESPONSE FROM MANUFACTURER

n/a

ADDTIONAL INFORMATION
SUPPORTING VISUALS/VIDEO
SUPPORTING DOCUMENT

OTHER PERFUSION
 REPORTS

London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Select and scroll to view chosen report
REPORT NO
002
1/1/10
Luer Cap Found In Venous Inlet Port
REPORT NO
001
1/1/09
Oxygenator Water To Blood Leak
Perfusion Report cards are shown in batches of 30 at any one time where available.
123
122
121
120
119
118
117
116
115
114
112
111
110
109
108
107
106
105
104
103
102
101
100
099
098
097
096
095
094
093
092
091
090
089
088
087
086
085
084
083
082
081
080
079
078
077
076
075
074
073
072
071
070
069
068
067
066
065
064
063
062
061
060
059
058
057
056
055
054
053
052
051
050
049
048
047
046
045
044
043
042
041
040
039
038
037
036
035
034
033
032
031
030
029
028
027
026
025
024
023
London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Know the report number? Scroll horizontally through the circle icons and select to view the appropriate report.