PERFUSION REPORTS

REPORT
083
DATE
1/6/2021
TITLE
IABP malfunction
DETAILS OF INCIDENT

The IABP's monitor became "white" and IABP therapy completely stopped while on a CICU patient around 20.00. The on-call perfusionist was called in. By the time the perfusionist arrived the IABP re-established its full functionality. All power cables and connections were found intact by both the CICU nurses and the on-call perfusionist. The perfusionist swapped the faulty IABP with another cardiosave and the device was placed out of service for investigations.

DETAILS OF INJURY TO PATIENT

The patients haemodynamics did not deteriorate during the 8minutes without IABP's support.

TYPE OF DEVICE/MANUFACTURER

Getinge Hybrid Cardiosave IABP

ACTION TAKEN

The device was taken out of service for investigation by Getinge. An incident report was completed and submitted to the hospital.

RESPONSE FROM MANUFACTURER

Please see attached.

ADDTIONAL INFORMATION
SUPPORTING VISUALS/VIDEO
SUPPORTING DOCUMENT

OTHER PERFUSION
 REPORTS

London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Select and scroll to view chosen report
REPORT NO
002
1/1/10
Luer Cap Found In Venous Inlet Port
REPORT NO
001
1/1/09
Oxygenator Water To Blood Leak
Perfusion Report cards are shown in batches of 30 at any one time where available.
123
122
121
120
119
118
117
116
115
114
112
111
110
109
108
107
106
105
104
103
102
101
100
099
098
097
096
095
094
093
092
091
090
089
088
087
086
085
084
083
082
081
080
079
078
077
076
075
074
073
072
071
070
069
068
067
066
065
064
063
062
061
060
059
058
057
056
055
054
053
052
051
050
049
048
047
046
045
044
043
042
041
040
039
038
037
036
035
034
033
032
031
030
029
028
027
026
025
024
023
London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Know the report number? Scroll horizontally through the circle icons and select to view the appropriate report.