TITLE
Electronic Gas Blender Fault
DETAILS OF INCIDENT

Routine CABG. Set up for case, all checks completed and satisfactory.

Harvesting of left radial artery and LIMA, so lines are not connected to the patient yet.

1045-Heparin given upon completion of left radial harvest.

1047-Connected lines up to the operating table.

1050-Checked gas flow through CDI, checked suckers and ACT reached 512

1050-1116-Connected to patient, gas flow set to 2.5 (this was enabled with no issues), RAP of arterial and venous lines.

1116-1117-Commenced CPB

1117-1122-Initial sight of arterial blood is red

Observation of venous and arterial blood darkening. Checked CDI and pO2 starting to drop. Perfusionist instructed nursing assistant to bring in another perfusionist. Perfusionist increases gas flow, no changes, no alarm from EGB.

Instructed anaesthetist to keep the lungs on. During this time tug tests of the gas line, no issues. Checked green gas line from EGB to oxygenator (including oxygen sensor); all in tact and no air leaks. Another perfusionist enters theatre and connects oxygen from a tank to the oxygenator and another perfusionist brings in a Sechrist gas blender.

Terminated CPB and transfused volume to the patient, patient stable.

1124-Perfusionist recommences CPB with the Sechrist gas blender. Case continues with no further problems and the patient is fine.

Post-CPB we connected the original EGB to the HLM and an error appears on the HLM "module/sensor defective". This had not occurred when the HLM was first turned on.

DETAILS OF INJURY TO PATIENT

None

TYPE OF DEVICE/MANUFACTURER

Liva Nova S5 Electronic Gas Blender

ACTION TAKEN

The faulty EGB was removed from circulation and the engineer from Liva Nova was informed. A datix was completed and all appropriate parties informed e.g. clinical technology. The Liva Nova rep subsequently sent the unit back to Germany for investigation.

RESPONSE FROM MANUFACTURER

Notes, Observations and Remarks:

EGB disinfected, opened, cleaned and checked.

Error confirmed, see report XXX

During pre-calibration error E1 was detected on O2.

Following parts were replaced, see items list.

Calibration was performed----OK

TSI and test run were performed----OK

ADDTIONAL INFORMATION
SUPPORTING VISUALS/VIDEO

See attached

SUPPORTING DOCUMENT

OTHER PERFUSION
 REPORTS

London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Select and scroll to view chosen report
REPORT NO
002
1/1/10
Luer Cap Found In Venous Inlet Port
REPORT NO
001
1/1/09
Oxygenator Water To Blood Leak
Perfusion Report cards are shown in batches of 30 at any one time where available.
123
122
121
120
119
118
117
116
115
114
112
111
110
109
108
107
106
105
104
103
102
101
100
099
098
097
096
095
094
093
092
091
090
089
088
087
086
085
084
083
082
081
080
079
078
077
076
075
074
073
072
071
070
069
068
067
066
065
064
063
062
061
060
059
058
057
056
055
054
053
052
051
050
049
048
047
046
045
044
043
042
041
040
039
038
037
036
035
034
033
032
031
030
029
028
027
026
025
024
023
London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Know the report number? Scroll horizontally through the circle icons and select to view the appropriate report.