On setting up for CPB and when testing the gas blender it was noted that there was an issue with the gas filter. This has happened with 2 filters out of 13 sets used with the same batch number.
The gas blender/ventilation module used was the Spectrum gas blender and when there is an occlusion of the gas line there is a visual representation of exclamation marks ie. !!!. Also the "Oxy-In" pressure readings were over 110mmHg which is outside the normal range and it causes a clicking sound in the ventilation module. This all occurred pre CPB before Diagnostics were switched on.
When Diagnostics are switched on there is a loud audible alarm and a flashing alert on the blender.
n/a
Liva Nova Gas Line Filter
The affected filters were switched out and replaced with a spare. The issue was rectified.
The filters were kept for return to manufacturer and a complaint filed.
Perfusion team were informed of the 2 faulty filters and must be vigilant when using the remaining sets from that batch.
Waiting for response