PERFUSION REPORTS

REPORT
103
DATE
25/7/2022
TITLE
Human Factors
DETAILS OF INCIDENT

Patient was primary CABG elective with normal BMI and no documented co-morbidities, allergies and sensitivities or relevant clinical history.

The Clinical Perfusion team was alerted to an elevated baseline ACT result (averaging 400 on re-testing).

The possibility of accidental infusion of Heparin was floated and discounted although a plain and Heparinase TEG test was started.

The on call consultant hematologist was consulted and they confirmed no evidence of coagulopathy with normal PT, PTT and INR.  There was no evidence of any exposure to anti coagulants in the pre op period on the ward and platelet inhibitors had all been stopped well before admission.

The TEG trace was inspected.  The Heparinase TEG showed an extended R time of 22 minutes, the plain test was still latent at 40 minutes, with no polymerization.

Accidental Heparin infusion was again suggested but the extended R time in the Heparinase TEG was held up as evidence that there was indeed an underlying coagulopathy.

Perfusionist, consultant Surgeon and Anesthetist discussed safety and possible way forward.  OP-CAB was discounted, retest ACT after administration of Protamine was also discounted.

Despite the seeming absence of a reason for the coagulopathy and extended ACT the team agreed that bypass with heparin would still be attempted.

Perfusion team set up pre and post membrane transducers which is not normal practice.  The operation proceeded with standard anticoagulation protocol.  Pre bypass ACT was >1000.  As a precaution the Perfusionist conducted a circuit RAP entirely removing the bypass prime from the circuit and recirculated for 5 mins.  After inspection of the membrane fibers and reservoir filters and final confirmation with the MDT, bypass was initiated.

The bypass proceeded uneventfully with trans-membrane pressures, and all bypass parameters within expected ranges.

The patient was successfully weaned from bypass and decannulated.  After Heparin reversal the ACT was measured and the result was 120.

DETAILS OF INJURY TO PATIENT

None

TYPE OF DEVICE/MANUFACTURER

N/A

ACTION TAKEN

A trust Ulysses incident report was submitted.  

The working assumption is that Heparin was accidentally infused instead of another pre induction Anesthetic drug.

RESPONSE FROM MANUFACTURER

N/A

ADDTIONAL INFORMATION
SUPPORTING VISUALS/VIDEO
SUPPORTING DOCUMENT

OTHER PERFUSION
 REPORTS

London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Select and scroll to view chosen report
REPORT NO
062
1/8/20
Human Factors Issue
REPORT NO
061
1/8/20
LIVA NOVA RESPONSE TO REPORTED CENTRIFUGAL HEAD ISSUE
REPORT NO
060
1/8/20
Oxygenator Fibre Leak
REPORT NO
059
1/5/20
Failure to remove CO2 effectively
REPORT NO
058
1/7/20
Centrifugal Head Issue 3
REPORT NO
057
1/6/20
Centrifugal Head Issue 2
REPORT NO
056
1/1/20
Centrifugal Head Issue
REPORT NO
055
1/2/20
Faulty Pressure Relief Valves
REPORT NO
054
1/4/20
Leaking ALF
REPORT NO
053
1/2/20
Arterial Line Filter Leak
REPORT NO
052
1/3/19
Gas Line Filter Issue
REPORT NO
051
1/12/19
Dual Lumen Cannula Movement
REPORT NO
050
1/10/19
Venous Cannula Damaged
REPORT NO
049
1/12/19
Oxygenation Issue
REPORT NO
048
1/8/19
Roller Pump Failure
REPORT NO
047
1/12/19
Broken Cannula Dilator
REPORT NO
046
6/12/19
Gas Blender Error
REPORT NO
045
20/11/19
Tubing Detachments
REPORT NO
044
27/11/19
Interruption to IABP counterpulsation
REPORT NO
043
1/7/19
Getinge Mal-Position Of IABP Catheter
REPORT NO
042
1/2/19
Avalon Dual Lumen Cannula Crack
REPORT NO
041
1/2/19
Custodial Rubber Fragmentation In Spiked Bag
REPORT NO
040
1/10/18
Reservoir Clot
REPORT NO
039
1/10/18
Pall Clot In Cardioplegia Circuit
REPORT NO
038
1/10/18
Maquet HL20 Cardioplegia Pump Error
REPORT NO
037
1/10/17
Temed D200 Haemoconcentrator Clot
REPORT NO
036
1/10/17
Oxygenator issue
REPORT NO
035
1/10/17
Medtronic Fusion Oxygenator Blood To Gas Leak
REPORT NO
034
Possible Medtronic Fusion Oxygenator Failure
REPORT NO
033
Fannin Heparin Ampules Shattering
Perfusion Report cards are shown in batches of 30 at any one time where available.
123
122
121
120
119
118
117
116
115
114
112
111
110
109
108
107
106
105
104
103
102
101
100
099
098
097
096
095
094
093
092
091
090
089
088
087
086
085
084
083
082
081
080
079
078
077
076
075
074
073
072
071
070
069
068
067
066
065
064
063
062
061
060
059
058
057
056
055
054
053
052
051
050
049
048
047
046
045
044
043
042
041
040
039
038
037
036
035
034
033
032
031
030
029
028
027
026
025
024
023
London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Know the report number? Scroll horizontally through the circle icons and select to view the appropriate report.