Hardware fault. Electronic gas Blender failure
DETAILS OF INCIDENT

Isolated AVR, low risk patient, only case of the day. (all times are approximate)

13.22 – Commencement of successful CPB, visual and numerical indicators were all in expected ranges, including colour of blood and saturations.

13.28 – Arterial blood colour started to change (darker) together with a gradual drop in venous saturation. Initial checks of circuit to identify any obvious issues to gas lines, including oxygen sensor, kink or detachment of gas line etc. On detachment of the gas line from the oxygenator there was no gas flow.

13.29 – Informed surgeon and anaesthetist that I was not oxygenating, and as the x-cross clamp had not been applied we came off bypass with the patient ventilated. Total time off bypass was approx. 4 mins. During this time the N+1 member of the department were asked to fetch and fit a Sechrist and the EGB was switched off, and a direct connection from the Sechrist to the oxygenator was established and isoflurane was omitted. The patient was sedated by the anaesthetic department medically. With a clear visual indication on the Sechrist blender that gas flow was established and flow could be felt post gas filter.

13.34 we commenced bypass for the second time. Second pump run was successful with no further oxygenation issues and the procedure concluded approx. 140 minutes later. Following the case the EGB was switched back on and a clear error code was displayed, E15, with an additional message on the dashboard screen reading “module/sensor defective”.

DETAILS OF INJURY TO PATIENT

The patient recovered well, with no cognitive dysfunction .

TYPE OF DEVICE/MANUFACTURER

Stockert S5 Electronic Gas Blender

ACTION TAKEN

The faulty EGB was removed from circulation and the engineer from LivaNova was informed. A Datix was completed and all appropriate parties informed e.g., clinical technology. The LivaNova rep subsequently sent the unit back to Germany for investigation. In addition, a request was made for the hospital estates team to test flow and pressure from the main hospital gas supply at the theatre pendants, as part of the full investigation to rule out issue locally.

RESPONSE FROM MANUFACTURER

Following the procedure, the unit was tested and an error code related to Module/Sensor defective appeared.

As per additional information AI-027472, Gas blender displayed the error code: E15 (Fault in AD transformer, An AD transformer fault has occurred).

This kind of malfunction can be caused by:

- improper hospital gas supply (a minimum pressure of 2 bar per connection should be observed. Indeed, input pressure too low can cause a discrepancy in the target and actual values, leading to the reported issue);

- a missed gas blender warm-up phase (user error);

- defective gas blender's component (A/D-converter);

- a defective DC/DC board.

The unit was manufactured in 2013 and no other similar events were complained.

The affected device was returned back to the manufacturer's site for repair. The device was opened, checked, and cleaned. The reported issue could not be reproduced and the device was found to be working within specifications. Subsequent functional verification testing was completed without further issues and the unit was returned to service.

During the tests performed after the first occurrence, no hardware malfunctions were detected, the unit was found working as expected and within specification.

During repair visit, front panel was replaced (reported in complaint 2023-01551).

Based on the above, it cannot be ruled out that a missed warm-up phase / an improper hospital gas supply have led to the reported issue.

No specific action was currently deemed necessary, LivaNova maintains and document periodic customer events monitoring process in order to evaluate actions for products improvement.

ADDTIONAL INFORMATION

The importance of N+1 is demonstrated in the response to the EGB failure and the ability to change out the defective equipment with a suitable back up. the incident further demonstrates the requirement for the availability of replacement equipment and the necessity of emergency drills and protocols.

SUPPORTING VISUALS/VIDEO
SUPPORTING DOCUMENT

OTHER PERFUSION
 REPORTS

London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Select and scroll to view chosen report
REPORT NO
092
1/10/21
Vapourizer Issue
REPORT NO
091
1/10/21
Oyxgenator Leak
REPORT NO
090
1/10/21
Breach of chemical cleaner
REPORT NO
089
1/9/21
Sechrist Gas Blender Sweep on ECMO
REPORT NO
088
1/9/21
Oxygenator High Pressure Events
REPORT NO
087
1/10/21
Leaking one way valve
REPORT NO
086
1/7/21
Electronic Gas Blender Fault
REPORT NO
085
1/2/21
Protamine Administration Issue
REPORT NO
084
1/6/21
Damaged roller pump lid
REPORT NO
083
1/6/21
IABP malfunction
REPORT NO
082
1/1/21
Accidental cardioplegia administration
REPORT NO
081
1/4/21
S5 Double Roller Pump incident
REPORT NO
080
1/5/21
Human factors issue
REPORT NO
079
1/6/21
Centrimag battery module failure
REPORT NO
078
1/6/21
Sampling manifold valve fault
REPORT NO
077
1/5/21
IABP transport issue
REPORT NO
076
1/4/21
Modified Ultrafiltration (MUF) Issue
REPORT NO
075
1/3/21
Electromagnetic interference with HLM
REPORT NO
074
1/2/21
HCU40 Failure
REPORT NO
073
1/2/21
HCU40 valve failures
REPORT NO
072
1/1/21
HCU40 failure
REPORT NO
071
1/1/21
LIVA NOVA STATEMENT ON CENTRIFUGAL HEAD
REPORT NO
070
1/1/21
Loose component on circuit
REPORT NO
069
1/12/20
Centrifugal Head noise
REPORT NO
068
1/11/20
Venous Reservoir disposable issue
REPORT NO
067
1/11/20
Failure of Heater Cooler due to leaking valves
REPORT NO
066
1/11/20
Failure of Valves on heater cooler unit
REPORT NO
065
1/9/20
Intra Aortic Balloon Pump (IABP) Failure
REPORT NO
064
1/9/20
Centrifugal head failure
REPORT NO
063
1/8/20
Centrifugal head issue 4
Perfusion Report cards are shown in batches of 30 at any one time where available.
123
122
121
120
119
118
117
116
115
114
112
111
110
109
108
107
106
105
104
103
102
101
100
099
098
097
096
095
094
093
092
091
090
089
088
087
086
085
084
083
082
081
080
079
078
077
076
075
074
073
072
071
070
069
068
067
066
065
064
063
062
061
060
059
058
057
056
055
054
053
052
051
050
049
048
047
046
045
044
043
042
041
040
039
038
037
036
035
034
033
032
031
030
029
028
027
026
025
024
023
London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Know the report number? Scroll horizontally through the circle icons and select to view the appropriate report.