PERFUSION REPORTS

REPORT
110
DATE
31/1/2023
TITLE
DISPOSABLE FAULT
DETAILS OF INCIDENT

    During the CPB, it was noticed  that the patient flow (as measured by the Spectrum M4 flow probe) was  significantly reduced (3.9 Lpm) comparative to pump flow at 6.0 Lpm. In  addition, the Spectrum emboli counter was increasing incrementally. At  first, this was thought to be due to flow probe positioning, but with further  investigation transpires that a kink had developed at the reservoir  outlet. 
 
 We concluded that the phenomena experienced was due to cavitation caused by  this kink, possibly resulting in mal-perfusion and/or micro-emboli. No injury  was caused to the patient.  

DETAILS OF INJURY TO PATIENT

No Injury

TYPE OF DEVICE/MANUFACTURER
ACTION TAKEN

A sling was used to ensure no further kinking of the tubing.

RESPONSE FROM MANUFACTURER

COMPLAINT INFORMATION

Complaint #:

2023-00505

Date of Event:

Jan 24, 2023

Hospital:

ABERDEEN ROYAL INFIRMARY

Event Description:

The pump boot outlet kink was noticed during aseptic assembly. An unsuccessful attempt was made to correct this defect, resulting in cutting out of this section of tubing (no pictures taken). The circuit was then deemed adequate for clinical use and CPB was commenced. During the CPB, it was noticed that the patient flow (as measured by the Spectrum M4 flow probe) was significantly reduced (3.9 Lpm) comparative to pump flow at 6.0 Lpm. In addition, the Spectrum emboli counter was increasing incrementally. At first, this was thought to be due to flow probe positioning, but with further investigation transpires that an additional kink had developed at the reservoir outlet (see picture attached). We concluded that the phenomena experienced was due to cavitation caused by this kink, possibly resulting in mal-perfusion and/or micro-emboli. This incident has been reported via the hospital adverse events system (DATIX) and I assume the MHRA will be informed. Luckily, no injury was caused to the patient.

PRODUCT INFORMATION:

Product Number:

IN01728H

Product Descripti on:

ABERDEEN AD INSPIRE 8F

Serial/LOT Number

: 2210270110

Investi gati on Summary:

LivaNova received a report stating that during set up, the Customer noted that the section of tubing #0191061 SG XS with a length of 60cm corresponding to the reservoir outlet in the pump loop was kinked. 1 unit of PTS PN IN01728H, lot 2210270110. No patient/user affected. Pictures of thissue were provided by the customer, confirming the kink. The sample was not available for investigation. Manufacturing records have been verified to be conforming to specifications. From complaints database review, three further similar events have been recorded on this PTS pack code (See #2022-00391, #2022-01109 and #2022-06709). No other complaint was found reported for the concerned batch of 24 units of this lot. Based on the above, the root cause could betraced back to two factors: - a tie-band that pressed against the tubing in question, which during sterilization caused a permanent mark or deformation. - a tubing length selected by the customer which should be changed in a following revision, since it is not long enough, this generates a tension in the tubing during set-up and the procedure, which can generate the kink due to the initial material deformation caused by the tie bands pressing on the tubing. As mitigation action, the drawing of this involved PTS will be updated to revision no. 03 in March 2023 in order to add specific pictures to help the operator in the assembly phase of this involved line in the table pack. The aim of this action is to reduce operator error as far as possible. Therefore, all lots manufactured before the new drawing revision release, are pre-mitigation action implementation. LivaNova maintains and documents periodic customer events monitoring process in order to evaluate actions for products improvement.

Root Causes:

Based on the above, the root cause could be traced back to two factors: - a tie-band that pressed against the tubing in question, which during sterilization caused a permanent mark or deformation. - a tubing length selected by the customer which should be changed in a following revision, since it is not long enough, this generates a tension in the tubing during set-up and the procedure, which can generate the kink due to the initial material deformation caused by the ti e bands pressing on the tubing.

Corrective Actions:

As mitigation action, the drawing of this involved PTS will be updated to revision no. 03 in March 2023 in order to add specific pictures to help the operator in the assembly phase of this involved line in the table pack. The aim of this action is to reduce operator error as far as possible. Therefore, all lots manufactured before the new drawing revision release, are pre-mitigation action implementation. LivaNova maintains and documents periodic customer events monitoring process in order to evaluate actions for products improvement.

Best regards,

Customer Quality-Team

Complaints Handling # 2023-00505

Kind regards,

Marti n

Marti n Brunt

Territory Sales Manager (CP & ATS)

Cardiac Surgery Division

United Kingdom

ADDTIONAL INFORMATION

It is interesting to note that as we monitor actual flow in the tube against the flow calculated by the roller pump it can show up this sort of problem highlighting the importance of inline monitoring.

 As the kink is from a pre-connected circuit that has suffered a imposed memory in the packaging and deformed on bypass.

Tubing length has now been altered to mitigate this problem.

SUPPORTING VISUALS/VIDEO
SUPPORTING DOCUMENT

OTHER PERFUSION
 REPORTS

London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Select and scroll to view chosen report
REPORT NO
032
Sucker Placed In Raceway In Wrong Direction
REPORT NO
031
High Transmembrane Pressure Gradients
REPORT NO
030
Faulty Connector On ECMO Soft Shell Reservoir Bag
REPORT NO
029
Sorin Inspire 8F Blood Leak From Temperature Port
REPORT NO
028
Sorin Inspire 8F Oxygenator Detached From Holding Arm
REPORT NO
027
Paragon Oxygenator Gas Inlet Leak
REPORT NO
026
Maquet IABP Console Failure
REPORT NO
025
Sorin EOS Gas To Blood Leak Whilst Priming
REPORT NO
024
Chalice Paragon Midi Heat Exchanger Leak Whilst Priming
REPORT NO
023
Chalice Paragon Heat Exchanger Water To Blood Leak
REPORT NO
022
Vent Pump Boot Tubing Inserted In Wrong Way Round
REPORT NO
021
Paragon Oxygenator Heat Exchanger Leak
REPORT NO
020
S3 E62 Pump Error
REPORT NO
019
Medtronic Potential Oxygenator Leak
REPORT NO
018
Medos Heat Exchanger Leak
REPORT NO
017
Avant D903 Has A Dual Chamber Reservoir Leak
REPORT NO
016
Additional Sucker Incorrectly Placed In Raceway
REPORT NO
015
Large Perished Tear Observed On Tubing At Back Of Vaporiser
REPORT NO
014
Gas Blender Flow Control Immobilised
REPORT NO
013
Isoflurane Vaporiser Seal Leak Due To Malposition
REPORT NO
012
1/2/13
Hypercoagulable Disorder May Have Led To Oxygenator Failure
REPORT NO
011
1/1/13
LV Vent Placed In Raceway In Wrong Direction
REPORT NO
010
1/3/12
Sucker Placed In Raceway In Wrong Direction
REPORT NO
009
1/2/12
Similarity In Heparin And St Thomas Cardioplegia Ampules
REPORT NO
008
1/7/11
Aorta Remained Clamped Post CPB
REPORT NO
007
1/5/11
Deficiencies In The QC At Terumo Ann Arbor, USA.
REPORT NO
006
1/4/11
Awareness Of Suckers When Using Bio-Glues
REPORT NO
005
1/2/11
Patient Exsanguinated From Additional Venous Line Post CPB
REPORT NO
004
1/1/11
Possible Inaccuracies In Colloid Research
REPORT NO
003
1/2/10
Difficulties In Visualising Blender Reading
Perfusion Report cards are shown in batches of 30 at any one time where available.
123
122
121
120
119
118
117
116
115
114
112
111
110
109
108
107
106
105
104
103
102
101
100
099
098
097
096
095
094
093
092
091
090
089
088
087
086
085
084
083
082
081
080
079
078
077
076
075
074
073
072
071
070
069
068
067
066
065
064
063
062
061
060
059
058
057
056
055
054
053
052
051
050
049
048
047
046
045
044
043
042
041
040
039
038
037
036
035
034
033
032
031
030
029
028
027
026
025
024
023
London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Know the report number? Scroll horizontally through the circle icons and select to view the appropriate report.