PERFUSION REPORTS

REPORT
112
DATE
20/2/2023
TITLE
HLM Hardware Error: S5 HLM - Dual roller cardioplegia
DETAILS OF INCIDENT

All pre-bypass checks complete. Bypass initiated, cross clamp applied and retrograde cardioplegia administration started. Unable to get sufficient flow rate due to coronary sinus pressure therefore stopped and moved to attempt antegrade cardioplegia via aortic root. When cardioplegia started the S5 console alarmed and displayed 'Pump malfunction' and it was noted that the crystalloid pump of the dual pump cardioplegia pump was not spinning.

Surgeon immediately informed that there was a serious failure and we were unable to deliver cardioplegia.

DETAILS OF INJURY TO PATIENT
TYPE OF DEVICE/MANUFACTURER

S5 Double Roller Pump85 / SN10E53864/ Date of Manufacture 2016-02-22

ACTION TAKEN

Surgeon immediately informed that there was a serious failure and we were unable to deliver cardioplegia. Cross clamp was removed and the heart was re-perfused. The cardioplegia pump was power cycled in isolation, testing was carried out and found that the issue had resolved. The decision was made to continue with the operation, which otherwise was uneventful.

RESPONSE FROM MANUFACTURER
ADDTIONAL INFORMATION

The safety committee has reviewed this incident and acknowledge that the older S5 pumps may have a trend with software errors that can account for these issues. Age and life span of the pump is something that must be monitored along with the incident rate and problems.

However, the pump was acquired in Feb 2016, so 7 years old.There have been no further issues with the pump. All pumps are regularlyserviced and maintained.

SUPPORTING VISUALS/VIDEO
SUPPORTING DOCUMENT

OTHER PERFUSION
 REPORTS

London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Select and scroll to view chosen report
REPORT NO
002
1/1/10
Luer Cap Found In Venous Inlet Port
REPORT NO
001
1/1/09
Oxygenator Water To Blood Leak
Perfusion Report cards are shown in batches of 30 at any one time where available.
123
122
121
120
119
118
117
116
115
114
112
111
110
109
108
107
106
105
104
103
102
101
100
099
098
097
096
095
094
093
092
091
090
089
088
087
086
085
084
083
082
081
080
079
078
077
076
075
074
073
072
071
070
069
068
067
066
065
064
063
062
061
060
059
058
057
056
055
054
053
052
051
050
049
048
047
046
045
044
043
042
041
040
039
038
037
036
035
034
033
032
031
030
029
028
027
026
025
024
023
London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Know the report number? Scroll horizontally through the circle icons and select to view the appropriate report.