PERFUSION REPORTS

REPORT
116
DATE
14/9/2023
TITLE
Disposable Issue: Reservoir in CPB circuit
DETAILS OF INCIDENT

Bypass circuit primed as per protocol, no problems observed and running for an hour before asked to divide lines. Noticed circuit pressure dropping and no flow on pump console although head was running. Checked all circuit couldn't find any problems. Asked for team member to assist and they noticed a pressure valve  blocking the outlet of the reservoir. This was an extra valve as the reservoir pressure valve was still in situ.

Changed out the reservoir to enable CPB to commence.

DETAILS OF INJURY TO PATIENT

No Harm

TYPE OF DEVICE/MANUFACTURER

Liva Nova: Product Code.: IN0168OH

ACTION TAKEN

Changed out the reservoir in first instance. Liva nova informed and perfusion sets with the same lot number isolated.

MHRA

C&V incident reporting

RESPONSE FROM MANUFACTURER

LivaNova received a report that an extra over/under valve was found to be loose inside HVR reservoir lot 2307030066 included into PTS Custom Pack code IN01680H lot 2306270086. There was no patient / user involvement.

Review of LivaNova complaints database revealed no other similar issue notified for batches concerned from the market.

Territory sales manager visited Customer's facility and provided visual evidence of the event that confirmed the following findings (#AI-029554):

- one over/under valve was correctly seated into its housing on the top lid of HVR as per the technical drawing of the device

- an additional valve was floating/movable inside the device

No physical return of the part was therefore arranged.

Through follow-up communication,LivaNova also learned that complained circuit was the first one opened by Customer and that other packs from affected batch number were rejected accordingly (#AI-031721).

Based on the manufacturing process of Inspire HVR Reservoir, the over/under valve is positioned on the dedicated housing in the top HVR lid with a press-tool. There is no possibility that reservoir is missing the valve since a leak test of the reservoir is performed at the end of the assembly line. Available video and pictures were shared with the LivaNova manufacturing department and the root cause was identified in an isolated manufacturing deviation: specifically, it is very likely that an operator poorly installed the over/under valve which dropped inside the unit in the further manipulation /handling steps of the device. At this point,a second operator reasonably did not intercept the floating material and re-processed the device by fitting a new over/under valve. Newly assembled unit successfully overcame the leak test and material passed undetected at visual checks before released (Inspire and PTS manufacturing floor). As per CP_MIR_SOP_000087, the personnel will be involved in a dedicated training meeting of customer complaint.

LivaNova maintains and documents periodic customer events monitoring process in order to evaluate actions for products improvement.

ADDTIONAL INFORMATION

The safety have reviewed the report and note that the N+1 enabled the team to respond to the emergency and appropriately identify the issue.

SUPPORTING VISUALS/VIDEO
SUPPORTING DOCUMENT

OTHER PERFUSION
 REPORTS

London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Select and scroll to view chosen report
REPORT NO
032
Sucker Placed In Raceway In Wrong Direction
REPORT NO
031
High Transmembrane Pressure Gradients
REPORT NO
030
Faulty Connector On ECMO Soft Shell Reservoir Bag
REPORT NO
029
Sorin Inspire 8F Blood Leak From Temperature Port
REPORT NO
028
Sorin Inspire 8F Oxygenator Detached From Holding Arm
REPORT NO
027
Paragon Oxygenator Gas Inlet Leak
REPORT NO
026
Maquet IABP Console Failure
REPORT NO
025
Sorin EOS Gas To Blood Leak Whilst Priming
REPORT NO
024
Chalice Paragon Midi Heat Exchanger Leak Whilst Priming
REPORT NO
023
Chalice Paragon Heat Exchanger Water To Blood Leak
REPORT NO
022
Vent Pump Boot Tubing Inserted In Wrong Way Round
REPORT NO
021
Paragon Oxygenator Heat Exchanger Leak
REPORT NO
020
S3 E62 Pump Error
REPORT NO
019
Medtronic Potential Oxygenator Leak
REPORT NO
018
Medos Heat Exchanger Leak
REPORT NO
017
Avant D903 Has A Dual Chamber Reservoir Leak
REPORT NO
016
Additional Sucker Incorrectly Placed In Raceway
REPORT NO
015
Large Perished Tear Observed On Tubing At Back Of Vaporiser
REPORT NO
014
Gas Blender Flow Control Immobilised
REPORT NO
013
Isoflurane Vaporiser Seal Leak Due To Malposition
REPORT NO
012
1/2/13
Hypercoagulable Disorder May Have Led To Oxygenator Failure
REPORT NO
011
1/1/13
LV Vent Placed In Raceway In Wrong Direction
REPORT NO
010
1/3/12
Sucker Placed In Raceway In Wrong Direction
REPORT NO
009
1/2/12
Similarity In Heparin And St Thomas Cardioplegia Ampules
REPORT NO
008
1/7/11
Aorta Remained Clamped Post CPB
REPORT NO
007
1/5/11
Deficiencies In The QC At Terumo Ann Arbor, USA.
REPORT NO
006
1/4/11
Awareness Of Suckers When Using Bio-Glues
REPORT NO
005
1/2/11
Patient Exsanguinated From Additional Venous Line Post CPB
REPORT NO
004
1/1/11
Possible Inaccuracies In Colloid Research
REPORT NO
003
1/2/10
Difficulties In Visualising Blender Reading
Perfusion Report cards are shown in batches of 30 at any one time where available.
123
122
121
120
119
118
117
116
115
114
112
111
110
109
108
107
106
105
104
103
102
101
100
099
098
097
096
095
094
093
092
091
090
089
088
087
086
085
084
083
082
081
080
079
078
077
076
075
074
073
072
071
070
069
068
067
066
065
064
063
062
061
060
059
058
057
056
055
054
053
052
051
050
049
048
047
046
045
044
043
042
041
040
039
038
037
036
035
034
033
032
031
030
029
028
027
026
025
024
023
London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Know the report number? Scroll horizontally through the circle icons and select to view the appropriate report.