PERFUSION REPORTS

REPORT
019
DATE
TITLE
Medtronic Potential Oxygenator Leak
DETAILS OF INCIDENT

Before bypass a small amount of fluid was noticed on the floor beneath the oxygenator, although this was believed to be condensation. On bypass a small amount of blood was noticed to be leaking from the oxygenator fiber block. The volume was minimal, approximately one drop every 5mins, the surgeon was informed and it was agreed to continue the case without further action unless the leak volume increased.

DETAILS OF INJURY TO PATIENT

None

TYPE OF DEVICE/MANUFACTURER

N/A

ACTION TAKEN

The case proceeded without any intervention. Following the procedure the oxygenator was flushed and retained for analysis by the supplier.

RESPONSE FROM MANUFACTURER

The subsequent report from Medtronic stated the following:

'At Medtronic’s Quality Laboratory, inspection showed a blood path through the mandrel-to-heat exchanger bond. No fiber leak was observed. The leak in this case was coming from a mechanical joint and clearly not a ‘water to blood’ leak across the heat exchanger. This was a leak path to the outside of the device due to a bonding issue that has since been corrected in production, you can see from the diagram where the leak path to the outside of the device is located and from where the Affinity in question was leaking.

Medtronic has identified that mandrel-heat exchanger bond leaks are most likely related to a UV adhesive void in the bond joint. In response to these leaks, an internal corrective action was initiated and resulted in improved manufacturing process. The mandrel is actually a mechanical joint between two pieces of plastic bonded together. It is the mating joint where the heat exchanger is attached to the oxygenator bundle, last step in the assembly process. The diagram shows where the joint (i.e. mandrel) is located. The joint is designed so that it is always under positive pressure and leaks from the inside out, thereby being a fluid leak from the blood side of the oxygenator to atmosphere.'

ADDTIONAL INFORMATION

N/A

SUPPORTING VISUALS/VIDEO
SUPPORTING DOCUMENT

OTHER PERFUSION
 REPORTS

London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Select and scroll to view chosen report
REPORT NO
062
1/8/20
Human Factors Issue
REPORT NO
061
1/8/20
LIVA NOVA RESPONSE TO REPORTED CENTRIFUGAL HEAD ISSUE
REPORT NO
060
1/8/20
Oxygenator Fibre Leak
REPORT NO
059
1/5/20
Failure to remove CO2 effectively
REPORT NO
058
1/7/20
Centrifugal Head Issue 3
REPORT NO
057
1/6/20
Centrifugal Head Issue 2
REPORT NO
056
1/1/20
Centrifugal Head Issue
REPORT NO
055
1/2/20
Faulty Pressure Relief Valves
REPORT NO
054
1/4/20
Leaking ALF
REPORT NO
053
1/2/20
Arterial Line Filter Leak
REPORT NO
052
1/3/19
Gas Line Filter Issue
REPORT NO
051
1/12/19
Dual Lumen Cannula Movement
REPORT NO
050
1/10/19
Venous Cannula Damaged
REPORT NO
049
1/12/19
Oxygenation Issue
REPORT NO
048
1/8/19
Roller Pump Failure
REPORT NO
047
1/12/19
Broken Cannula Dilator
REPORT NO
046
6/12/19
Gas Blender Error
REPORT NO
045
20/11/19
Tubing Detachments
REPORT NO
044
27/11/19
Interruption to IABP counterpulsation
REPORT NO
043
1/7/19
Getinge Mal-Position Of IABP Catheter
REPORT NO
042
1/2/19
Avalon Dual Lumen Cannula Crack
REPORT NO
041
1/2/19
Custodial Rubber Fragmentation In Spiked Bag
REPORT NO
040
1/10/18
Reservoir Clot
REPORT NO
039
1/10/18
Pall Clot In Cardioplegia Circuit
REPORT NO
038
1/10/18
Maquet HL20 Cardioplegia Pump Error
REPORT NO
037
1/10/17
Temed D200 Haemoconcentrator Clot
REPORT NO
036
1/10/17
Oxygenator issue
REPORT NO
035
1/10/17
Medtronic Fusion Oxygenator Blood To Gas Leak
REPORT NO
034
Possible Medtronic Fusion Oxygenator Failure
REPORT NO
033
Fannin Heparin Ampules Shattering
Perfusion Report cards are shown in batches of 30 at any one time where available.
123
122
121
120
119
118
117
116
115
114
112
111
110
109
108
107
106
105
104
103
102
101
100
099
098
097
096
095
094
093
092
091
090
089
088
087
086
085
084
083
082
081
080
079
078
077
076
075
074
073
072
071
070
069
068
067
066
065
064
063
062
061
060
059
058
057
056
055
054
053
052
051
050
049
048
047
046
045
044
043
042
041
040
039
038
037
036
035
034
033
032
031
030
029
028
027
026
025
024
023
London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Know the report number? Scroll horizontally through the circle icons and select to view the appropriate report.