PERFUSION REPORTS

REPORT
019
DATE
TITLE
Medtronic Potential Oxygenator Leak
DETAILS OF INCIDENT

Before bypass a small amount of fluid was noticed on the floor beneath the oxygenator, although this was believed to be condensation. On bypass a small amount of blood was noticed to be leaking from the oxygenator fiber block. The volume was minimal, approximately one drop every 5mins, the surgeon was informed and it was agreed to continue the case without further action unless the leak volume increased.

DETAILS OF INJURY TO PATIENT

None

TYPE OF DEVICE/MANUFACTURER

N/A

ACTION TAKEN

The case proceeded without any intervention. Following the procedure the oxygenator was flushed and retained for analysis by the supplier.

RESPONSE FROM MANUFACTURER

The subsequent report from Medtronic stated the following:

'At Medtronic’s Quality Laboratory, inspection showed a blood path through the mandrel-to-heat exchanger bond. No fiber leak was observed. The leak in this case was coming from a mechanical joint and clearly not a ‘water to blood’ leak across the heat exchanger. This was a leak path to the outside of the device due to a bonding issue that has since been corrected in production, you can see from the diagram where the leak path to the outside of the device is located and from where the Affinity in question was leaking.

Medtronic has identified that mandrel-heat exchanger bond leaks are most likely related to a UV adhesive void in the bond joint. In response to these leaks, an internal corrective action was initiated and resulted in improved manufacturing process. The mandrel is actually a mechanical joint between two pieces of plastic bonded together. It is the mating joint where the heat exchanger is attached to the oxygenator bundle, last step in the assembly process. The diagram shows where the joint (i.e. mandrel) is located. The joint is designed so that it is always under positive pressure and leaks from the inside out, thereby being a fluid leak from the blood side of the oxygenator to atmosphere.'

ADDTIONAL INFORMATION

N/A

SUPPORTING VISUALS/VIDEO
SUPPORTING DOCUMENT

OTHER PERFUSION
 REPORTS

London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Select and scroll to view chosen report
REPORT NO
032
Sucker Placed In Raceway In Wrong Direction
REPORT NO
031
High Transmembrane Pressure Gradients
REPORT NO
030
Faulty Connector On ECMO Soft Shell Reservoir Bag
REPORT NO
029
Sorin Inspire 8F Blood Leak From Temperature Port
REPORT NO
028
Sorin Inspire 8F Oxygenator Detached From Holding Arm
REPORT NO
027
Paragon Oxygenator Gas Inlet Leak
REPORT NO
026
Maquet IABP Console Failure
REPORT NO
025
Sorin EOS Gas To Blood Leak Whilst Priming
REPORT NO
024
Chalice Paragon Midi Heat Exchanger Leak Whilst Priming
REPORT NO
023
Chalice Paragon Heat Exchanger Water To Blood Leak
REPORT NO
022
Vent Pump Boot Tubing Inserted In Wrong Way Round
REPORT NO
021
Paragon Oxygenator Heat Exchanger Leak
REPORT NO
020
S3 E62 Pump Error
REPORT NO
019
Medtronic Potential Oxygenator Leak
REPORT NO
018
Medos Heat Exchanger Leak
REPORT NO
017
Avant D903 Has A Dual Chamber Reservoir Leak
REPORT NO
016
Additional Sucker Incorrectly Placed In Raceway
REPORT NO
015
Large Perished Tear Observed On Tubing At Back Of Vaporiser
REPORT NO
014
Gas Blender Flow Control Immobilised
REPORT NO
013
Isoflurane Vaporiser Seal Leak Due To Malposition
REPORT NO
012
1/2/13
Hypercoagulable Disorder May Have Led To Oxygenator Failure
REPORT NO
011
1/1/13
LV Vent Placed In Raceway In Wrong Direction
REPORT NO
010
1/3/12
Sucker Placed In Raceway In Wrong Direction
REPORT NO
009
1/2/12
Similarity In Heparin And St Thomas Cardioplegia Ampules
REPORT NO
008
1/7/11
Aorta Remained Clamped Post CPB
REPORT NO
007
1/5/11
Deficiencies In The QC At Terumo Ann Arbor, USA.
REPORT NO
006
1/4/11
Awareness Of Suckers When Using Bio-Glues
REPORT NO
005
1/2/11
Patient Exsanguinated From Additional Venous Line Post CPB
REPORT NO
004
1/1/11
Possible Inaccuracies In Colloid Research
REPORT NO
003
1/2/10
Difficulties In Visualising Blender Reading
Perfusion Report cards are shown in batches of 30 at any one time where available.
123
122
121
120
119
118
117
116
115
114
112
111
110
109
108
107
106
105
104
103
102
101
100
099
098
097
096
095
094
093
092
091
090
089
088
087
086
085
084
083
082
081
080
079
078
077
076
075
074
073
072
071
070
069
068
067
066
065
064
063
062
061
060
059
058
057
056
055
054
053
052
051
050
049
048
047
046
045
044
043
042
041
040
039
038
037
036
035
034
033
032
031
030
029
028
027
026
025
024
023
London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Know the report number? Scroll horizontally through the circle icons and select to view the appropriate report.