PERFUSION REPORTS

REPORT
088
DATE
1/9/2021
TITLE
Oxygenator High Pressure Events
DETAILS OF INCIDENT

During use of the Affinity Fusion Oxygenators high pre-membrane pressures were observed. Fifteen minutes into bypass and whilst cooling to 32degrees, the pre-membrane pressures of 380mmHg were climbing to 500mmHg. The target patient blood flow was 5.6LPM (relatively large patient). The pressure isolators and equipment were checked. As pressures were continuing to rise a second opinion was sought and the decision was taken to change out the oxygenator.

At that stage (20mins in to bypass, maximum revolutions per minute on the centrifugal pump (3500RPM) could only maintain a flow to the patient of 3.5LPM approximately. Blood gas was performed and the patients parameters were still okay. The patient was not yet cross clamped and cardioplegia had not been given. The anaesthetist ventilated the patient while the perfusionist changed out the oxygenator (approximately 5 minutes to changeout the oxygenator, done in a controlled manner as the patient was stable).

Upon initiating bypass with a new (second) oxygenator, high pre-membrane oxygenator pressures were noted. The high pressure alarm was set off at 550mmHg. A flow of approximately 5.2LPM was maintained and the blood gas results of the patient were acceptable. The second period of bypass was around 2hours long. As the patient was being rewarmed toward the the end of bypass, the pre-membrane pressure started to falling to normal levels (330mmHg pre and 210mmHg post membrane). During both bypass periods the ACT's were good (>480 on Medtronic ACT plus).

A HMS Hepcon heparin management system was used. This patient had a target heparin concentration of 3mg/kg. Despite ACT's being good the HMS results stated to provide more heparin (anti-coagulate the patient more). During the bypass an additional 20,000units of heparin were given. The patient's Hb was 150g/liter, but the perfusionist did try and run at slightly lower of 120g/liter. There were no known patient blood anomalies prior to bypass. The patient was given Aprotinin and additional Aprotinin at the end of bypass (prescribed - as had passed the allergy check).

This is one of 8 recent HPE with this oxygenator this year (2021). We have been using them for a number of years without similar problems prior to these events.

DETAILS OF INJURY TO PATIENT

N/A

TYPE OF DEVICE/MANUFACTURER

Medtronic Affinity Fusion Oxygenator

ACTION TAKEN

Oxygenator collected by supplier and investigated.

RESPONSE FROM MANUFACTURER
ADDTIONAL INFORMATION
SUPPORTING VISUALS/VIDEO
SUPPORTING DOCUMENT

OTHER PERFUSION
 REPORTS

London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Select and scroll to view chosen report
REPORT NO
062
1/8/20
Human Factors Issue
REPORT NO
061
1/8/20
LIVA NOVA RESPONSE TO REPORTED CENTRIFUGAL HEAD ISSUE
REPORT NO
060
1/8/20
Oxygenator Fibre Leak
REPORT NO
059
1/5/20
Failure to remove CO2 effectively
REPORT NO
058
1/7/20
Centrifugal Head Issue 3
REPORT NO
057
1/6/20
Centrifugal Head Issue 2
REPORT NO
056
1/1/20
Centrifugal Head Issue
REPORT NO
055
1/2/20
Faulty Pressure Relief Valves
REPORT NO
054
1/4/20
Leaking ALF
REPORT NO
053
1/2/20
Arterial Line Filter Leak
REPORT NO
052
1/3/19
Gas Line Filter Issue
REPORT NO
051
1/12/19
Dual Lumen Cannula Movement
REPORT NO
050
1/10/19
Venous Cannula Damaged
REPORT NO
049
1/12/19
Oxygenation Issue
REPORT NO
048
1/8/19
Roller Pump Failure
REPORT NO
047
1/12/19
Broken Cannula Dilator
REPORT NO
046
6/12/19
Gas Blender Error
REPORT NO
045
20/11/19
Tubing Detachments
REPORT NO
044
27/11/19
Interruption to IABP counterpulsation
REPORT NO
043
1/7/19
Getinge Mal-Position Of IABP Catheter
REPORT NO
042
1/2/19
Avalon Dual Lumen Cannula Crack
REPORT NO
041
1/2/19
Custodial Rubber Fragmentation In Spiked Bag
REPORT NO
040
1/10/18
Reservoir Clot
REPORT NO
039
1/10/18
Pall Clot In Cardioplegia Circuit
REPORT NO
038
1/10/18
Maquet HL20 Cardioplegia Pump Error
REPORT NO
037
1/10/17
Temed D200 Haemoconcentrator Clot
REPORT NO
036
1/10/17
Oxygenator issue
REPORT NO
035
1/10/17
Medtronic Fusion Oxygenator Blood To Gas Leak
REPORT NO
034
Possible Medtronic Fusion Oxygenator Failure
REPORT NO
033
Fannin Heparin Ampules Shattering
Perfusion Report cards are shown in batches of 30 at any one time where available.
123
122
121
120
119
118
117
116
115
114
112
111
110
109
108
107
106
105
104
103
102
101
100
099
098
097
096
095
094
093
092
091
090
089
088
087
086
085
084
083
082
081
080
079
078
077
076
075
074
073
072
071
070
069
068
067
066
065
064
063
062
061
060
059
058
057
056
055
054
053
052
051
050
049
048
047
046
045
044
043
042
041
040
039
038
037
036
035
034
033
032
031
030
029
028
027
026
025
024
023
London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Know the report number? Scroll horizontally through the circle icons and select to view the appropriate report.