PERFUSION REPORTS

REPORT
088
DATE
1/9/2021
TITLE
Oxygenator High Pressure Events
DETAILS OF INCIDENT

During use of the Affinity Fusion Oxygenators high pre-membrane pressures were observed. Fifteen minutes into bypass and whilst cooling to 32degrees, the pre-membrane pressures of 380mmHg were climbing to 500mmHg. The target patient blood flow was 5.6LPM (relatively large patient). The pressure isolators and equipment were checked. As pressures were continuing to rise a second opinion was sought and the decision was taken to change out the oxygenator.

At that stage (20mins in to bypass, maximum revolutions per minute on the centrifugal pump (3500RPM) could only maintain a flow to the patient of 3.5LPM approximately. Blood gas was performed and the patients parameters were still okay. The patient was not yet cross clamped and cardioplegia had not been given. The anaesthetist ventilated the patient while the perfusionist changed out the oxygenator (approximately 5 minutes to changeout the oxygenator, done in a controlled manner as the patient was stable).

Upon initiating bypass with a new (second) oxygenator, high pre-membrane oxygenator pressures were noted. The high pressure alarm was set off at 550mmHg. A flow of approximately 5.2LPM was maintained and the blood gas results of the patient were acceptable. The second period of bypass was around 2hours long. As the patient was being rewarmed toward the the end of bypass, the pre-membrane pressure started to falling to normal levels (330mmHg pre and 210mmHg post membrane). During both bypass periods the ACT's were good (>480 on Medtronic ACT plus).

A HMS Hepcon heparin management system was used. This patient had a target heparin concentration of 3mg/kg. Despite ACT's being good the HMS results stated to provide more heparin (anti-coagulate the patient more). During the bypass an additional 20,000units of heparin were given. The patient's Hb was 150g/liter, but the perfusionist did try and run at slightly lower of 120g/liter. There were no known patient blood anomalies prior to bypass. The patient was given Aprotinin and additional Aprotinin at the end of bypass (prescribed - as had passed the allergy check).

This is one of 8 recent HPE with this oxygenator this year (2021). We have been using them for a number of years without similar problems prior to these events.

DETAILS OF INJURY TO PATIENT

N/A

TYPE OF DEVICE/MANUFACTURER

Medtronic Affinity Fusion Oxygenator

ACTION TAKEN

Oxygenator collected by supplier and investigated.

RESPONSE FROM MANUFACTURER
ADDTIONAL INFORMATION
SUPPORTING VISUALS/VIDEO
SUPPORTING DOCUMENT

OTHER PERFUSION
 REPORTS

London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Select and scroll to view chosen report
REPORT NO
032
Sucker Placed In Raceway In Wrong Direction
REPORT NO
031
High Transmembrane Pressure Gradients
REPORT NO
030
Faulty Connector On ECMO Soft Shell Reservoir Bag
REPORT NO
029
Sorin Inspire 8F Blood Leak From Temperature Port
REPORT NO
028
Sorin Inspire 8F Oxygenator Detached From Holding Arm
REPORT NO
027
Paragon Oxygenator Gas Inlet Leak
REPORT NO
026
Maquet IABP Console Failure
REPORT NO
025
Sorin EOS Gas To Blood Leak Whilst Priming
REPORT NO
024
Chalice Paragon Midi Heat Exchanger Leak Whilst Priming
REPORT NO
023
Chalice Paragon Heat Exchanger Water To Blood Leak
REPORT NO
022
Vent Pump Boot Tubing Inserted In Wrong Way Round
REPORT NO
021
Paragon Oxygenator Heat Exchanger Leak
REPORT NO
020
S3 E62 Pump Error
REPORT NO
019
Medtronic Potential Oxygenator Leak
REPORT NO
018
Medos Heat Exchanger Leak
REPORT NO
017
Avant D903 Has A Dual Chamber Reservoir Leak
REPORT NO
016
Additional Sucker Incorrectly Placed In Raceway
REPORT NO
015
Large Perished Tear Observed On Tubing At Back Of Vaporiser
REPORT NO
014
Gas Blender Flow Control Immobilised
REPORT NO
013
Isoflurane Vaporiser Seal Leak Due To Malposition
REPORT NO
012
1/2/13
Hypercoagulable Disorder May Have Led To Oxygenator Failure
REPORT NO
011
1/1/13
LV Vent Placed In Raceway In Wrong Direction
REPORT NO
010
1/3/12
Sucker Placed In Raceway In Wrong Direction
REPORT NO
009
1/2/12
Similarity In Heparin And St Thomas Cardioplegia Ampules
REPORT NO
008
1/7/11
Aorta Remained Clamped Post CPB
REPORT NO
007
1/5/11
Deficiencies In The QC At Terumo Ann Arbor, USA.
REPORT NO
006
1/4/11
Awareness Of Suckers When Using Bio-Glues
REPORT NO
005
1/2/11
Patient Exsanguinated From Additional Venous Line Post CPB
REPORT NO
004
1/1/11
Possible Inaccuracies In Colloid Research
REPORT NO
003
1/2/10
Difficulties In Visualising Blender Reading
Perfusion Report cards are shown in batches of 30 at any one time where available.
123
122
121
120
119
118
117
116
115
114
112
111
110
109
108
107
106
105
104
103
102
101
100
099
098
097
096
095
094
093
092
091
090
089
088
087
086
085
084
083
082
081
080
079
078
077
076
075
074
073
072
071
070
069
068
067
066
065
064
063
062
061
060
059
058
057
056
055
054
053
052
051
050
049
048
047
046
045
044
043
042
041
040
039
038
037
036
035
034
033
032
031
030
029
028
027
026
025
024
023
London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Know the report number? Scroll horizontally through the circle icons and select to view the appropriate report.